8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
If the Fault Is Intermittent
Repeat Test Function
If the failure is intermittent, do the following:
1. To switch on the repeat function, press the following:
SERVICE MENU
System
2. Then press the following:
RETURN
TESTS
3. Select the test desired and press the following:
EXECUTE TEST
4. Press
to stop the function. The test repeat function is explained in
Preset
"Service Key Menus and Error Messages."
GPIB Failures
If you have performed
there is a GPIB problem in the analyzer, replace the CPU board (A7).
Service Guide
TEST OPTIONS
"Step 3. GPIB Systems Check" on page 4-10
Digital Control Troubleshooting
If the Fault Is Intermittent
REPEAT ON
and you are certain that
Chapter 10 ,
6- 19