Serial Number Correction Constants (Test 49) - Agilent Technologies 8719ET Service Manual

Network analyzer
Hide thumbs Also See for 8719ET:
Table of Contents

Advertisement

8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers

Serial Number Correction Constants (Test 49)

This procedure stores the analyzer serial number in the A7 CPU assembly EEPROMs.
CAUTION
Perform this procedure only if the A7 CPU assembly has been replaced.
Analyzer warmup time: None.
1. Make sure the A7 Switch is in the alter position (ALT).
2. Record the ten character serial number that is on the analyzer rear panel identification
label.
The serial number may also be entered using the optional external keyboard.
NOTE
Plug the keyboard into the "DIN KYBD" connector on the rear panel and
press
Preset
3. To erase the HP/Agilent logo, press the following:
PRESET: FACTORY
Preset
Display
MORE
4. Enter the serial number by rotating the front panel knob to position the arrow below
each character of the instrument serial number, and then pressing
to enter each character. Enter a total of ten characters: two letters and eight digits.
Press
if you make a mistake.
5. Press
when you have finished entering the title. Double check that the correct
DONE
serial number appears in the title area. If you made a mistake at this point, return to
step 3.
CAUTION
You cannot correct mistakes after you perform step 6, unless you contact the
nearest Agilent Technologies sales or service center, for a clear serial number
keyword. Then you must perform
Constants (Test 50)"
6. Press
SERVICE MENU
System
7. Observe the analyzer for the results of the routine:
If the analyzer displays the message Serial Cor DONE, you have completed this
procedure.
8. If no more correction constant routines are going to be performed, return the A7 Switch
to the NRM position and perform the
Service Guide
.
Preset
TITLE
ERASE TITLE
and repeat this procedure.
TESTS
"EEPROM Backup Disk Procedure," on page
Adjustments and Correction Constants
Serial Number Correction Constants (Test 49)
"Protected Option Numbers Correction
EXECUTE TEST
49
x1
SELECT LETTER
.
YES
3-33.
3- 11

Advertisement