Service Key Menus and Error Messages
Service Related Error Messages in Alphabetical Order
INITIALIZATION FAILED
Error Number
The disk initialization failed, probably because the disk is damaged.
47
INSUFFICIENT MEMORY, PWR MTR CAL OFF
Error Number
There is not enough memory space for the power meter calibration array.
154
Increase the available memory by clearing one or more save/recall registers, or
by reducing the number of points.
NO CALIBRATION CURRENTLY IN PROGRESS
Error Number
The
69
already in progress. Start a new calibration.
NO FAIL FOUND
Service Error
The self-diagnose function of the instrument operates on an internal test failure.
Number 114
At this time, no failure has been detected.
NO FILE(S) FOUND ON DISK
Error Number
No files of the type created by an analyzer store operation were found on the
45
disk. If you requested a specific file title, that file was not found on the disk.
NO IF FOUND: CHECK R INPUT LEVEL
Error Number
The first IF signal was not detected during pretune. Check the front panel R
5
channel jumper. If there is no visible problem with the jumper, refer to
7 , "Source Troubleshooting."
NO SPACE FOR NEW CAL.CLEAR REGISTERS
Error Number
You cannot store a calibration set due to insufficient memory. You can free more
70
memory by clearing a saved instrument state from an internal register (which
may also delete an associated calibration set, if all the instrument states using
the calibration kit have been deleted). You can store the saved instrument state
and calibration set to a disk before clearing them. After deleting the instrument
states, press
NOT ALLOWED DURING POWER METER CAL
Error Number
When the analyzer is performing a power meter calibration, the GPIB bus is
198
unavailable for other functions such as printing or plotting.
10-36
RESUME CAL SEQUENCE
Preset
to run the memory packer.