8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
8719ET/ES and 8720ET/ES Models
1. Refer to
Table 2-10
for the particular analyzer under test. After you enter the value, press
2. Wait for the analyzer to complete a full sweep.
DATA → MEMORY
3. Press
Display
∆ MODE MENU
4. Press
Marker
.
0
x1
Recording P
Data
1
5. Press
Sweep Setup
6. Enter the power value that is listed in the P
analyzer under test. After you enter the value, press
7. Wait for the analyzer to complete a full sweep.
8. Press
Marker Search
9. Read the marker value from the analyzer display and add 5 dB. Record the calculated
value in the "P
Max." column of the "Source Linearity" performance test record.
1
10.Press
SEARCH: MIN
11.Read the marker value from the analyzer display and add 5 dB. Record the calculated
value in the "P
Min." column of the "Source Linearity" performance test record.
1
Recording P
Data
2
12.Repeat steps 5 through 11, except in step 6 enter the power value in the P
step 11 read the marker and add 10 dB to arrive at the calculated value. Record the
calculated value in the "P
performance test record.
Recording P
Data
3
13.Repeat steps 5 through 11, except in step 6 enter the power value in the P
step 11 read the marker and subtract 5 dB to arrive at the calculated value. Record the
calculated value in the "P
performance test record.
Recording P
Data
4
14.Repeat steps 5 through 11, except in step 6 enter the power value in the P
step 11 read the marker and subtract 10 dB to arrive at the calculated value. Record the
calculated value in the "P
performance test record.
Service Guide
and set the analyzer to the power value that is listed in the P
DATA/MEM
FIXED MKR POSITION
.
POWER
.
SEARCH: MAX
.
Max." and "P
2
Max." and "P
3
Max." and "P
4
System Verification and Performance Tests
.
row of
Table 2-10
1
x1
.
Min." columns of the "Source Linearity"
2
Min." columns of the "Source Linearity"
3
Min." columns of the "Source Linearity"
4
Performance Tests
Ref
.
x1
FIXED MKR VALUE
for the particular
row and in
2
row and in
3
row and in
4
row
2- 39

Advertisement

Table of Contents
loading

This manual is also suitable for:

8720et8722et8719es8720es8722es

Table of Contents