8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
LIMITS[NORM/SPCL]
(POWLLIST) accesses the following Edit List menu to allow
POWER LOSS
LOSS/SENSR LISTS
•
USE SENSOR A/B
list for use in power meter calibration measurements.
•
CAL FACTOR SENSOR A
to allow modification of the calibration data table for power sensor A.
•
CAL FACTOR SENSOR B
to allow modification of the calibration data table for power sensor B.
•
POWER LOSS
modification of the external power loss data table that corrects
coupled-arm power loss when a directional coupler samples the RF
output.
DUMP GRAPH
Edit List Menu
To access this menu, press
LOSS/SENSR LISTS
CAL FACTOR SENSOR B
SEGMENT
(SEDI[D])
EDIT
(SDEL)
DELETE
(SADD)
ADD
(CLEL)
CLEAR LIST
(EDITDONE)
DONE
Service Guide
selects either normal (NORM) or Special (SPCL) (tighter)
limits for the Operator's Check. The SPCL limits are
useful for a guard band.
modification of the external power loss data table.
accesses the power loss/sensor lists menu:
selects the A or B power sensor calibration factor
(POWLLIST) accesses the Edit List menu to allow
generates printed graphs of verification results when
activated during a system verification.
SERVICE MENU
System
and then press one of the following:
or
POWER LOSS
selects a segment (frequency point) to be edited, deleted
from, or added to the current data table. Works with the
entry controls.
allows modification of frequency, cal factor and loss values
previously entered in the current data table.
deletes frequency, cal factor and loss values previously
entered in the current data table.
adds new frequency, cal factor and loss values to the
current data table up to a maximum of 12 segments
(frequency points, PTS).
deletes the entire current data table (or list) when
pressed. Press
returns to the previous menu.
Service Key Menus and Error Messages
Service Menus - Internal Diagnostics
(CALFSENA) accesses the Edit List menu
(CALFSENB) accesses the Edit List menu
TEST OPTIONS
CAL FACTOR SENSOR A
.
to avoid deletion.
NO
or
is
YES
10- 7