Check For A Faulty Assembly - Agilent Technologies 8719ET Service Manual

Network analyzer
Hide thumbs Also See for 8719ET:
Table of Contents

Advertisement

8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
Figure 5-6 A15W1 Plug Detail

Check for a Faulty Assembly

This procedure checks for a faulty assembly that might be shutting down the
preregulator (A15) via one of the following lines (refer to
• A15W1 connecting to the post regulator (A8)
• the +5VCPU line through the motherboard
• the +5VDIG line through the motherboard
Do the following:
1. Switch off the analyzer.
2. Ensure that A15W1 is reconnected to A8 (refer to
NOTE
Always switch off the analyzer before removing or disconnecting assemblies.
To identify specific cables and assemblies that are not shown in this chapter,
refer to
When extensive disassembly is required, refer to
Replacement and Post-Repair Procedures."
Service Guide
Chapter 13 , "Replaceable Parts."
Power Supply Troubleshooting
If the Red LED of the Preregulator (A15) Is On
Figure
5-1).
Figure
5-5).
Chapter 14 , "Assembly
5- 11