System Verification and Performance Tests
Preliminary Tests
Preliminary Tests
CAUTION
Use an antistatic work surface and wrist strap to lessen the chance of
electrostatic discharge.
Required Equipment and Tools
Equipment Description
Calibration Kit
RF Cable Set
Verification Kit
Check the Temperature and Humidity
1. Measure the temperature and humidity of the environment and record the values in
Table 2-14, "Performance Test Record for All Models," on page
specified at an ambient temperature of +23 ±3 °C. Therefore, the environmental
temperature must remain in the range of +20 °C to +26 °C. Once the measurement
calibration has been done, the ambient temperature must be held to range of ± 1 °C.
2. Open the calibration and verification kits and place all the devices on top of the foam so
they will reach room temperature. The temperature of the devices is important because
device dimensions and therefore the electrical characteristics change with temperature.
3. Switch on the power to the analyzer.
NOTE
To achieve the maximum system stability, allow the analyzer to warm up for
at least 30 minutes.
2-6
8719/20
85052B
85056A
85131F
85133F
85053B
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
8722
85057B
2-46. The performance is
Service Guide