7. If you have difficulty with the printer, check the following:
• If the interface on your printer is GPIB, verify that the printer address is set to 701
(or change the setting in the analyzer to match the printer).
• If the interface on your printer is serial or parallel, be sure that you selected the
printer port and the printer type correctly (refer to the user's guide for more
information on how to perform these tasks).
8. Press
Cal
CAL KIT
9. Press the following:
SERVICE MENU
System
TESTS SYS VER TESTS
EXECUTE TEST
10.The analyzer displays Sys Ver Init DONE when the initialization procedure is
complete.
CAUTION
At this point, do not preset the analyzer or recall another instrument state.
You must use the instrument state from this procedure for the next
procedure.
Service Guide
and select the type of calibration kit used.
SELECT CAL KIT
System Verification and Performance Tests
System Verification