8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
Table 11-4 8722ET/ES Uncorrected Port Characteristics
Parameter and Option
Directivity
1
(ES and ET models)
Source Match
(ES Standard, Option 400)
Source Match (ET Models)
Source Match
(ES Option 007, Option 085)
Load Match
(ES Standard, Option 400)
Load Match
(ES Option 007, Option 085)
Load Match (ET Models)
Reflection Tracking
2
(ES Models)
Reflection Tracking
2
(ET models)
Transmission Tracking
2
(ES Models)
Transmission Tracking
1
(ET Models)
Crosstalk (ES Models)
Crosstalk (ET Models)
1. Does not include the effect of the cable set on the test ports.
2. Excludes roll-off below 500 MHz, which is typically −18 dB at 100 MHz, and −25 dB at
50 MHz.
3. Measurement Conditions: normalized to a thru, measured with two shorts, 10 Hz IF
bandwidth, averaging factor of 8, alternate mode, source power set to the lesser of maximum
power out or the maximum receiver power.
4. Limited by noise floor.
Service Guide
0.05 to 0.5 GHz
23 dB
17 dB
1
1
16 dB
17 dB
1
18 dB
1
21 dB
1
22 dB
1
±3 dB
±3 dB
±3 dB
±3 dB
3
4
60 dB
3
95 dB
Frequency Range
.5 to 2 Ghz
2 to 8 GHz
23 dB
21 dB
17 dB
12 dB
20 dB
14 dB
17 dB
15 dB
18 dB
15 dB
21 dB
17 dB
20 dB
20 dB
±3 dB
±3 dB
±3 dB
±3 dB
±3 dB
±3 dB
±3 dB
±3 dB
85 dB
85 dB
95 dB
95 dB
Error Terms
Error Term Inspection
8 to 20 GHz
20 to 40 GHz
16 dB
15 dB
11 dB
7 dB
11 dB
10 dB
11 dB
8 dB
12 dB
10 dB
13 dB
10 dB
20 dB
15 dB
±3 dB
±6 dB
±3 dB
±4 dB
±3 dB
±6 dB
±3 dB
±3 dB
82 dB
72 dB
88 dB
82 dB