Start Troubleshooting Here
Step 3. GPIB Systems Check

Troubleshooting Systems with Multiple Peripherals

Connect any other system peripherals (but not a controller) to the analyzer one at a time
and check their functionality. Any problems observed are in the peripherals, cables, or are
address problems (see previous section).

Troubleshooting Systems with Controllers

Passing the preceding checks indicates that the analyzer's peripheral functions are
normal. Therefore, if the analyzer has not been operating properly with an external
controller, check the following:
• The GPIB interface hardware is incorrectly installed or not operational. (See the
"Operating Concepts" chapter in the user's guide.)
• The programming syntax is incorrect. (Refer to your programmer's guide.)
If the analyzer appears to be operating unexpectedly but has not completely failed, go to
"Step 4. Faulty Group Isolation" on page
4-12
4-13.