Group Sensor Data (Ch1 To Ch2); Function Group Configuration - Endress+Hauser Proline Prosonic Flow 93C Description Of Device Functions

Profibus dp/pa ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93C:
Table of Contents

Advertisement

Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA
Endress+Hauser
8.4

Group SENSOR DATA (CH1 to CH2)

8.4.1

Function group CONFIGURATION

BASIC FUNCTION
G
BASIC FUNCTION → SENSOR DATA (CH1 to CH2) → CONFIGURATION
K-FACTOR
Use this function to view the current calibration factor for the measuring tube and the
(6800)
measuring sensors.
Display:
5-digit floating-point number, (including sign)
Factory setting:
Depends on the measuring tube and the calibration.
ZERO POINT
Use this function to view the zero-point correction value for the measuring tube and the
(6803)
measuring sensors The calibration at the factory determines the zero-point correction
value.
Display:
Max. 5-digit number
Factory setting:
Depends on the measuring tube and the calibration
C0
Use this function to view the current correction factor of the sound velocity for the
(6806)
measuring tube and the measuring sensors. The correction factor is determined during
factory calibration.
Display:
Max. 5-digit number
Factory setting:
1.0000 (= no correction)
PROFIBUS DP
GBA
(depends on PROFIBUS module)
PROFIBUS PA
GCA
PROCESS PARAMETER
GIA
CH1
PROCESS PARAMETER
GIB
CH2
SYSTEM PARAMETER
GLA
CH1
SYSTEM PARAMETER
GLB
CH2
SENSOR DATA CH1
GNA
SENSOR DATA CH2
GNB
Functional description
8 Block BASIC FUNCTION
CONFIGURATION
CONFIGURATION
680
680
103