Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual
Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual

Endress+Hauser Proline Prosonic Flow 93T Portable Operating Instructions Manual

Portable ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93T Portable:

Advertisement

Brief Operating Instructions
Proline Prosonic Flow 93T Portable
Portable ultrasonic flow measuring system
These Brief Operating Instructions are not intended to replace the
Operating Instructions provided in the scope of supply.
Detailed information is provided in the Operating Instructions and
the additional documentation on the CD-ROM supplied.
The complete device documentation consists of:
• These Brief Operating Instructions
• Depending on the device version:
– Operating Instructions and the Description of Device Functions
– Approvals and safety certificates
– Special safety instructions in accordance with the approvals for
the device (e.g. explosion protection, pressure equipment
directive, etc.)
– Additional device-specific information
KA00034D/06/EN/13.11
71136632

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Prosonic Flow 93T Portable

  • Page 1 Brief Operating Instructions Proline Prosonic Flow 93T Portable Portable ultrasonic flow measuring system These Brief Operating Instructions are not intended to replace the Operating Instructions provided in the scope of supply. Detailed information is provided in the Operating Instructions and the additional documentation on the CD-ROM supplied.
  • Page 2 Table of contents Proline Prosonic Flow 93T Portable Table of contents 1 Safety instructions ........3 1.1 Designated use .
  • Page 3: Safety Instructions

    Proline Prosonic Flow 93T Portable Safety instructions Safety instructions Designated use • The measuring device described in these Operating Instructions is to be used only for measuring the flow rate of liquids in closed pipes, e.g.: – Acids, alkalis, paints, oils –...
  • Page 4: Operational Safety

    If the temperature of the fluid is high, implement sufficient measures to prevent burning or scalding. • Endress+Hauser will be happy to assist in clarifying any questions on approvals, their application and implementation. Safety conventions Warning! “Warning”...
  • Page 5: Installation

    Proline Prosonic Flow 93T Portable Installation Installation Installation conditions 2.1.1 Dimensions For the dimensions of the measuring device, see the associated Technical Information on the CD-ROM. Mounting location Correct flow measurement is possible only if the pipe is full. Entrained air or gas forming in the pipe can result in an increase in measuring errors.
  • Page 6 Installation Proline Prosonic Flow 93T Portable Orientation Vertical Recommended orientation with upward direction of flow (View A). With this orientation, entrained solids will sink and gases will rise away from the sensor when the fluid is stagnant. The piping can be completely drained and protected against solids buildup.
  • Page 7 Proline Prosonic Flow 93T Portable Installation Inlet and outlet run If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the following inlet and outlet runs is recommended in order to ensure measuring accuracy.
  • Page 8: Preparatory Steps Prior To Installation

    Installation Proline Prosonic Flow 93T Portable Preparatory steps prior to installation Depending on the conditions specific to the measuring point (e.g. Clamp On, number of traverses, fluid, etc.), a number of preparatory steps have to be taken before actually installing...
  • Page 9: Determining The Installation Distances

    Proline Prosonic Flow 93T Portable Installation Determining the installation distances 2.5.1 Running the "Sensor Installation" Quick Setup menu Calling the Quick Setup All the installation distances required for sensor mounting are determined with the Quick Setup. Note! If you are not familiar with the operation of the device → ä 28.
  • Page 10 Installation Proline Prosonic Flow 93T Portable Running the Quick Setup for Clamp On type of mounting • Only the steps necessary for the type of mounting (Clamp On) are described. • Enter or select installation-specific values or the values specified here.
  • Page 11: Mechanical Preparation

    Proline Prosonic Flow 93T Portable Installation Mechanical preparation The way in which the sensors are secured differs on account of the pipe nominal diameter and the sensor type. Overview of possible ways to secure the various sensors: Sensor For the measuring range...
  • Page 12 Installation Proline Prosonic Flow 93T Portable Guide the sensor holder over the pipe. A0013542 Fig. 5: Setting the sensor holder on the pipe Release the quick release of the retaining bracket (a) and push the retaining bracket up against the pipe.
  • Page 13 Proline Prosonic Flow 93T Portable Installation Fix the sensor holder in place by: – Tightening the quick release of the retaining bracket (a) – Tightening the quick release (b) A0013544 Fig. 7: Fixing the sensor holder in place Quick release of retaining bracket...
  • Page 14 Installation Proline Prosonic Flow 93T Portable 2.6.2 Premounting the strapping bands (medium nominal diameters) When mounting on a pipe with a nominal diameter of DN ≤ 200 (8"). Sensor: Prosonic Flow P (DN 50 to 4000 / 2 to 160") First strapping band Fit the mounting bolt over the strapping band.
  • Page 15 Proline Prosonic Flow 93T Portable Installation 2.6.3 Premounting the strapping bands (large nominal diameters) When mounting on a pipe with a nominal diameter of DN > 200 (8"). For sensor: Prosonic Flow P (DN 50 to 4000 / 2 to 160") Measure the pipe circumference.
  • Page 16 Installation Proline Prosonic Flow 93T Portable 2.6.4 Mounting with strapping bands (flexible) For sensor Prosonic Flow P (DN 50 to 4000 / 2 to 160") " Caution! • Each time you use the strapping bands, check that the ratchet locks and springs function safely beforehand.
  • Page 17: Installing Prosonic Flow P (Dn 15 To 65 / ½ To 2½")

    Proline Prosonic Flow 93T Portable Installation Installing Prosonic Flow P (DN 15 to 65 / ½ to 2½") 2.7.1 Mounting the sensor Prerequisites • The sensor holder is already mounted → ä 11. • The distance of the sensor holder is set (sensor distance) → ä 11.
  • Page 18 Installation Proline Prosonic Flow 93T Portable Mount as illustrated in the graphic (steps 1 to 5): A0013551 Fig. 12: Mounting the sensors This completes the mounting process. The sensors can now be connected to the transmitter via the connecting cables → ä 25.
  • Page 19: Installing Prosonic Flow P (Dn 50 To 4000 / 2 To 160") (Clamp On)

    Proline Prosonic Flow 93T Portable Installation Installing Prosonic Flow P (DN 50 to 4000 / 2 to 160") (Clamp On) Note! Sensor orientation shown in the following sketches is for visual purposes only. Please apply the recommended orientation → ä 6.
  • Page 20 Installation Proline Prosonic Flow 93T Portable With the second measuring wire: – Fit the cable lug over the mounting bolt of the strapping band that is already securely mounted. – Run the measuring wire counterclockwise around the pipe. – Fit the fixer over the mounting bolt of the strapping band that can still be moved.
  • Page 21 Proline Prosonic Flow 93T Portable Installation Coat the contact surfaces of the sensors with an even layer of coupling fluid approx. 1 mm (0.04") thick, going from the groove through the center to the opposite edge. A0011373 Fig. 16: Coating the contact surfaces of the sensor with coupling fluid Insert the sensor into the sensor holder.
  • Page 22 Installation Proline Prosonic Flow 93T Portable 2.8.2 Installation for measurement via two traverses Prerequisites • The installation distance (position sensor) is known → ä 8. • The strapping bands are already mounted → ä 14/→ ä 15. Material The following material is needed for mounting: •...
  • Page 23 Proline Prosonic Flow 93T Portable Installation Tighten the strapping band so that it cannot slip. Remove the mounting rail from the mounting bolt. Fit the sensor holders over the individual mounting bolts and tighten securely with the retaining nut. A0001117 Fig.
  • Page 24: Post-Installation Check

    Installation Proline Prosonic Flow 93T Portable Screw the connecting cable into the individual sensor. A0001112 Fig. 21: Mounting the sensor and connecting the connecting cable This completes the mounting process. The sensors can now be connected to the transmitter via the connecting cables →...
  • Page 25: Wiring

    Proline Prosonic Flow 93T Portable Wiring Wiring Charging the NiMH storage battery Warning! • The storage battery for the device (NiMH storage batteries) may only be charged with the charger supplied. Other equipment could cause the battery to overheat. • Compare the information on the nameplate of the charger with the local supply voltage and frequency.
  • Page 26: Connecting The Connecting Cable

    Connecting the connecting cable Warning! • Only use the connecting cables supplied by Endress+Hauser. • The measuring device only complies with the general safety requirements in accordance with EN 61010-1 and the EMC requirements of IEC/EN 61326 during storage battery operation.
  • Page 27: Commissioning

    Proline Prosonic Flow 93T Portable Commissioning Commissioning Switching on the measuring device The measuring device is switched on by pressing the ON/OFF switch ≥ 3 seconds (→ ä 26, No. 1). After being switched on, the measuring system performs a number of internal test functions. As...
  • Page 28: Operation

    Commissioning Proline Prosonic Flow 93T Portable Operation 4.2.1 Display elements Display lines/fields Main line for primary measured values Additional line for additional measured +24.502 variables/status variables Information line for bar graph display for example +1863.97 Info icons, e.g. volume flow Current measured values –50...
  • Page 29: Navigating Within The Function Matrix

    Proline Prosonic Flow 93T Portable Commissioning Navigating within the function matrix > 3 s – – – – – – – – A0007665 F → Enter the function matrix (starting with measured value display) P → Select the Block (e.g. USER INTERFACE) F →...
  • Page 30: Calling The Commissioning Quick Setup

    Commissioning Proline Prosonic Flow 93T Portable Calling the Commissioning Quick Setup All the functions needed for commissioning are called up automatically with the Quick Setup. The functions can be changed and adapted to suit the individual processes. F → Enter the function matrix (starting with measured value display) P →...
  • Page 31 Proline Prosonic Flow 93T Portable Commissioning Endress+Hauser...
  • Page 32 KA00034D/06/EN/13.11 71136632 FM+SGML 6.0...