Group System (System Ch2); Function Group Configuration - Endress+Hauser Proline Prosonic Flow 93C Description Of Device Functions

Profibus dp/pa ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93C:
Table of Contents

Advertisement

9 Block SUPERVISION
110
9.1

Group SYSTEM (SYSTEM CH2)

9.1.1

Function group CONFIGURATION

SUPERVISION
J
ALARM DELAY
Use this function to specify a time period for suppressing the appearance of fault or notice
(8005)
messages.
This suppression acts on:
• Display
• Current output
• Frequency output
• Relay output
• PROFIBUS DP/PA
User input:
0 to 100 s (in steps of one second)
Factory setting:
0 s
"
If this function is activated, error and notice messages are delayed by the time
corresponding to the setting before being forwarded to the higher-order controller
(process controller, etc.).
It is therefore imperative to check in advance in order to make sure whether a delay of
this nature could affect the safety requirements of the process. If error and notice
messages cannot be suppressed, a value of 0 seconds must be entered here.
PERMANENT STORAGE
This function displays whether permanent storage of all parameters in the EEPROM is
(8007)
switched on or off.
Options:
OFF
ON
Factory setting:
ON
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA
SYSTEM
Functional description
SUPERVISION → SYSTEM → CONFIGURATION
Caution!
JAA
CONFIGURATION
Endress+Hauser
800

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline Prosonic Flow 93C and is the answer not in the manual?

Questions and answers

Table of Contents