Group SYSTEM PARAMETER (CH1 to CH2)

8.3.1

Function group CONFIGURATION

BASIC FUNCTION
G
BASIC FUNCTION → SYSTEM PARAMETER (CH1...CH2) → CONFIGURATION
INSTALLATION
Use this function to reverse the sign of the flow quantity, if necessary.
DIRECTION SENSOR
(6600)
Options:
NORMAL
INVERSE (REVERSE)
Factory setting:
NORMAL
MEASURING MODE
Use this function to specify the flow direction for outputting the signal:
(6601)
Unidirectional:
Signal is only output if the flow direction is positive (forward flow). Flow in the negative
flow direction (reverse flow) is not taken into account or totalized by the measuring
system.
Bidirectional:
Signal is output with flow in both directions (forward and reverse flow).
Options:
UNIDIRECTIONAL
BIDIRECTIONAL
Factory setting:
UNIDIRECTIONAL
FLOW DAMPING
!
(6603)
The system damping acts on all functions and outputs of the measuring device.
Use this function to set the filter depth of the digital filter. This reduces the sensitivity of
the measuring signal to interference peaks (e.g. high solids content, gas bubbles in the
fluid, etc.). The system reaction time increases with the filter setting.
User input:
0 to 100 s
Factory setting:
0 s
PROFIBUS-DP
GBA
(depends on PROFIBUS module)
PROFIBUS-PA
GCA
PROCESS PARAMETER
GIA
CH1
PROCESS PARAMETER
GIB
CH2
SYSTEM PARAMETER
GLA
CH1
SYSTEM PARAMETER
GLB
CH2
Functional description
Note!
8 Block BASIC FUNCTION
CONFIGURATION
CONFIGURATION
660
660
101