Device Functions Proline Promag 55 FOUNDATION Fieldbus
Endress+Hauser
7.4

Group SENSOR DATA

7.4.1

Function group CONFIGURATION

BASIC FUNCTION
G
BASIC FUNCTION
All sensor data (calibration factors, zero (point) and nominal diameter) are set at the factory and saved on the S-DAT
sensor memory chip.
!
Note!
The individual values of the functions are also provided on the sensor nameplate.
"
Caution!
Under normal circumstances you should not change the following parameter settings, because changes affect numerous
functions of the entire measuring facility in general and the accuracy of the measuring system in particular. For this
reason, the functions described below cannot be changed even when you enter your private code.
Contact the Endress+Hauser service organization if you have any questions about these functions.
K-FACTOR POSITIVE
Use this function to display the actual calibration factor (positive flow direction) for the
(6801)
sensor. The calibration factor is determined and set at the factory.
User interface:
5-digit fixed point number: 0.5000 to 2.0000
Factory setting:
Depends on nominal diameter and calibration
K-FACTOR NEGATIVE
Use this function to display the actual calibration factor (negative flow direction) for the
(6802)
sensor. The calibration factor is determined and set at the factory.
User interface:
5-digit fixed point number: 0.5000 to 2.0000
Factory setting:
Depends on nominal diameter and calibration
ZERO POINT
Use this function to view the actual zero-point correction value for the sensor.
(6803)
Zero-point correction is determined and set at the factory.
User interface:
Max. 4-digit number: –1000 to +1000
Factory setting:
Depends on nominal diameter and calibration
NOMINAL DIAMETER
Use this function to view the nominal diameter of the sensor. The nominal diameter
(6804)
depends on the size of the sensor and is set at the factory.
User interface:
2 to 2000 mm or 1/12 to 78"
Factory setting:
Depends on the size of the sensor
FOUNDATION FIELDBUS GGA
PROCESS PARAMETER
GIA
SYSTEM PARAMETER
GLA
SENSOR DATA
GNA
Function description
SENSOR DATA
7 Block BASIC FUNCTION
CONFIGURATION
CONFIGURATION
680
51