8 Block BASIC FUNCTION
90
8.1.3

Function group TOTALIZER

BASIC FUNCTION
G
BASIC FUNCTION → PROFIBUS-DP / PROFIBUS-PA → TOTALIZER
SELECT TOTALIZER
Use this function to select a totalizer.
(6130)
Options:
TOTALIZER 1
TOTALIZER 2
TOTALIZER 3
Factory setting:
TOTALIZER 1
!
If the option PROFILE-GSD was selected in the SELECTION GSD function (6140) the
only option available in this function is TOTALIZER 1.
TOTALIZER OUT VALUE
Use this function to display the OUT value incl. the unit and status of the totalizer
(6131)
selected in the SELECT TOTALIZER function (6130).
OVERFLOW
Use this function to view the totaled overflow for the totalizer aggregated since
(6132)
measuring commenced.
Total flow quantity is represented by a floating-point number consisting of max. 7 digits.
This function can be used to view higher numerical values (>9,999,999) as overflows.
The effective quantity is thus the total of OVERFLOW plus the value returned by the
SUM function.
Example:
Reading for 2 overflows: 2 10
The value displayed in the SUM function = 196,845.7 m
Effective total quantity = 20,196,845.7 m
Display:
Integer with exponent, including sign (e.g. 2 E
CHANNEL
Use this function to assign a measured variable to the totalizer.
(6133)
Options:
OFF
VOLUME FLOW AVERAGE
Factory setting:
VOLUME FLOW AVERAGE
!
The totalizer is reset to "0" as soon as the selection is changed.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA
PROFIBUS-DP
GBA
(depends on PROFIBUS module)
PROFIBUS-PA
GCA
Functional description
Note!
3
7
m
(= 20,000,000 m
Note!
CONFIGURATION
FUNCTION BLOCKS
TOTALIZER
3
).
3
3
7
)