Endress+Hauser Proline Prosonic Flow 93C Description Of Device Functions page 91

Profibus dp/pa ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93C:
Table of Contents

Advertisement

Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA
Endress+Hauser
BASIC FUNCTION → PROFIBUS-DP / PROFIBUS-PA → TOTALIZER
UNIT TOTALIZER
Use this function to define a unit for the measured variable of the totalizer.
(6134)
Options:
Metric → cm
US → cc; af; ft
bbl (filling tanks)
Imperial → gal; Mgal; bbl (beer); bbl (petrochemicals)
Factory setting:
3
m
SET TOTALIZER
Use this function to assign a status to the totalizer.
(6135)
Options:
TOTALIZE
Total the measured variable selected in the CHANNEL function (6133).
RESET
Reset the totalizer to zero.
PRESET
The totalizer is set to the value defined in the PRESET TOTALIZER function (6136).
!
Note that selecting RESET or PRESET resets the totalizer to "0" or sets it to the preset
value respectively, but does not stop the totalizer. This means that it immediately
recommences totaling from the new setting. If you want to stop the totalizer you must
select HOLD in the TOTALIZER MODE (6137) function.
Factory setting:
TOTALIZE
PRESET TOTALIZER
Use this function to define a start value for the totalizer.
(6136)
!
This value is not accepted by the totalizer unless the PRESET option is selected in the
SET TOTALIZER function (6135).
User input:
–99999...99999
Factory setting:
0
Functional description
3
3
3
; dm
; m
; ml; l; hl; Ml
3
; oz f; gal; Mgal; bbl (normal fluids); bbl (beer); bbl (petrochemicals);
Note!
Note!
8 Block BASIC FUNCTION
91

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline Prosonic Flow 93C and is the answer not in the manual?

Questions and answers

Table of Contents