Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 1 Using this Manual Using this Manual There are various ways of locating the description of a function of your choice in the manual: Using the table of contents to locate a function description The designations of all the cells in the function matrix are listed in the table of contents.
2 Function matrix Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Function matrix General layout of the function matrix The function matrix consists of four levels: Blocks -> Groups -> Function groups -> Functions 0000 0001 0002 0003 0009 …...
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 2 Function matrix 2.1.5 Codes identifying cells Each cell (block, group, function group and function) in the function matrix has an individual, unique code. Blocks: The code is a letter (A, B, C, etc.) Groups: The code consists of three letters (AAA, ABA, BAA, etc.).
2 Function matrix Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Max. number of writes The number of writes to the EEPROM is technically restricted to a maximum of 1 million. Attention must be paid to this limit since, if exceeded, it results in data loss and measuring device failure. For...
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 2 Function matrix Function matrix Prosonic Flow 93C PROFIBUS BLOCKS GROUPS FUNCTION GROUPS → → MEASURED VARIABLES MEASURING VALUES Page 11 → (s. Page 10) SYSTEM UNITS Page 13 ↓ → →...
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 3 Block MEASURED VARIABLES Group MEASURING VALUES 3.1.1 Function group MAIN VALUES CH1 → → MEASURED VARIABL. A MEASURING VALUES MAIN VALUES CH1 Functional description MEASURED VARIABLES → MEASURING VALUES → MAIN VALUES CH1 SIGNAL STRENGTH CH1 The signal strength appears on the display (channel 1).
3 Block MEASURED VARIABLES Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 3.1.3 Function group CALCULATED MAIN VALUES → → MEASURED VARIABL. A MEASURING VALUES CALC. MAIN VALUES Functional description MEASURED VARIABLES → MEASURING VALUES → CALC. MAIN VALUES The calculated measured values appear on the display. The measured values of both channels are used when calculating the values.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 3 Block MEASURED VARIABLES Group SYSTEM UNITS 3.2.1 Function group CONFIGURATION → MEASURED VARIABL. A MEASURING VALUES ↓ → SYSTEM UNITS CONFIGURATION Functional description MEASURED VARIABLES → SYSTEM UNITS → CONFIGURATION You can select the units for measured variables in this function group.
Page 14
3 Block MEASURED VARIABLES Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description MEASURED VARIABLES → SYSTEM UNITS → CONFIGURATION UNIT VOLUME Use this function to select the unit for displaying the volume. (0403) Options: Metric: Cubic centimeter → cm³...
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 3 Block MEASURED VARIABLES 3.2.2 Function group ADDITIONAL CONFIGURATION → MEASURED VARIABL. A MEASURING VALUES ↓ → SYSTEM UNITS CONFIGURATION ↓ ADD. CONFIGURATION 042 Functional description MEASURED VARIABLES → SYSTEM UNITS → ADDITIONAL CONFIGURATION UNIT TEMPERATURE Use this function to select the unit for displaying the fluid temperature.
Page 16
3 Block MEASURED VARIABLES Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description MEASURED VARIABLES → SYSTEM UNITS → ADDITIONAL CONFIGURATION FORMAT DATE/TIME Use this function to select the date and time format of the calibration history. (0429) Options: DD.MM.YY 24 H...
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 4 Block QUICK-SETUP Block QUICK-SETUP Function Block Group Functions groups QUICK-SETUP T-DAT → → → COMMISSION. SAVE/LOAD (1002) P. 17 (1009) P. 17 Functional description QUICK-SETUP SETUP Use this function to start the Quick Setup menu for commissioning.
4 Block QUICK-SETUP Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Quick Setup "Commissioning" In the case of measuring devices without a local display, the individual parameters and functions must be configured via the operating program, e.g. FieldCare. If the measuring device is equipped with a local display, all the important device parameters for standard operation, as well as additional functions, can be configured quickly and easily by means of the following Quick Setup menus.
Page 19
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 4 Block QUICK-SETUP Note! • The display returns to the function SETUP COMMISSIONING (1002) if you press the ESC key combination during parameter interrogation. • If you answer YES to the question regarding the "Automatic configuration of the display", the...
"ENGLISH". • You can change the language group via the configuration software FieldCare. Please do not hesitate to contact your Endress+Hauser sales office if you have any questions. DISPLAY DAMPING Use this function to enter a time constant defining how the display reacts to severely...
Page 22
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description USER INTERFACE → CONTROL → BASIC CONFIGURATION CONTRAST LCD Use this function to optimize display contrast to suit local operating conditions. (2003) User input: 10…100% Factory setting: BACKLIGHT Use this function to optimize the backlight to suit local operating conditions.
HOME position. • You can also disable programming in this function by entering any number (other than the defined private code). • The Endress+Hauser service organization can be of assistance if you mislay your personal code. DEF. PRIVATE CODE...
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 5.1.3 Function group OPERATION → → USER INTERFACE CONTROL BASIC CONFIGURATION 200 ↓ UN-/LOCKING ↓ OPERATION Functional description USER INTERFACE → CONTROL → OPERATION TEST DISPLAY Use this function to test the operability of the local display and its pixels.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 5 Block USER INTERFACE Group MAIN LINE 5.2.1 Function group CONFIGURATION → USER INTERFACE CONTROL ↓ → MAIN LINE CONFIGURATION Functional description USER INTERFACE → MAIN LINE → CONFIGURATION 1 = Main line...
Page 26
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description USER INTERFACE → MAIN LINE → CONFIGURATION FORMAT Use this function to define the maximum number of places after the decimal point (2202) displayed for the reading in the main line.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 5 Block USER INTERFACE 5.2.2 Function group MULTIPLEX → USER INTERFACE CONTROL ↓ → MAIN LINE CONFIGURATION ↓ MULTIPLEX Functional description USER INTERFACE → MAIN LINE → MULTIPLEX ASSIGN Use this function to define a second reading to be displayed in the main line alternatively (2220) (every 10 seconds) with the reading defined in the ASSIGN function (2200).
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Group ADDITION LINE 5.3.1 Function group CONFIGURATION → USER INTERFACE CONTROL ↓ MAIN LINE ↓ → ADDITION LINE CONFIGURATION Functional description USER INTERFACE → ADDITION LINE → CONFIGURATION...
Page 29
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 5 Block USER INTERFACE Functional description USER INTERFACE → ADDITION LINE → CONFIGURATION FORMAT Note! (2402) This function is not available unless a number was selected in the ASSIGN function (2400). Use this function to define the maximum number of places after the decimal point displayed for the reading in the additional line.
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 5.3.2 Function group MULTIPLEX → USER INTERFACE CONTROL ↓ MAIN LINE ↓ → ADDITION LINE CONFIGURATION ↓ MULTIPLEX Functional description USER INTERFACE → ADDITION LINE → MULTIPLEX ASSIGN...
Page 31
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 5 Block USER INTERFACE Functional description USER INTERFACE → ADDITION LINE → MULTIPLEX FORMAT Note! (2422) This function is not available unless a number was selected in the ASSIGN function (2420). Use this function to define the maximum number of places after the decimal point for the second value displayed in the additional line.
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Group INFORMATION LINE 5.4.1 Function group CONFIGURATION → USER INTERFACE CONTROL ↓ MAIN LINE ↓ ADDITION LINE ↓ → INFORMATION LINE CONFIGURATION Functional description USER INTERFACE → INFORMATION LINE → CONFIGURATION...
Page 33
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 5 Block USER INTERFACE Functional description USER INTERFACE → INFORMATION LINE → CONFIGURATION 100% VALUE Note! (2601) This function is not available unless one of the following was selected in the ASSIGN function (2400): •...
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 5.4.2 Function group MULTIPLEX → USER INTERFACE CONTROL ↓ MAIN LINE ↓ ADDITION LINE ↓ → INFORMATION LINE CONFIGURATION ↓ MULTIPLEX Functional description USER INTERFACE → INFORMATION LINE → MULTIPLEX...
Page 35
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 5 Block USER INTERFACE Functional description USER INTERFACE → INFORMATION LINE → MULTIPLEX FORMAT Note! (2622) This function is not available unless a number was selected in the ASSIGN function (2600). Use this function to define the maximum number of places after the decimal point for the second value displayed in the information line.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Group CURRENT OUTPUT 1 6.1.1 Function group CONFIGURATION → → OUTPUTS CURRENT OUTPUT 1 CONFIGURATION Functional description OUTPUTS → CURRENT OUTPUT 1 → CONFIGURATION (only with PROFIBUS DP) ASSIGN CURRENT Use this function to assign a measured variable to the current output.
Page 38
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → CURRENT OUTPUT 1 → CONFIGURATION (only with PROFIBUS DP) CURRENT SPAN Use this function to define the current span. The selection specifies the operational range (4001) and the lower and upper signal on alarm.
Page 39
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → CURRENT OUTPUT 1 → CONFIGURATION (only with PROFIBUS DP) VALUE 0_4 mA Use this function to assign the 0/4 mA current a value. (4002) The value can be greater or less than the value assigned to 20 mA, (function VALUE 20 mA (4003)).
Page 40
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → CURRENT OUTPUT 1 → CONFIGURATION (only with PROFIBUS DP) VALUE 0_4 mA Parameter setting example A: (Continued) 1. VALUE 0_4 mA (4002) = not equal to zero flow (e.g. –5 m VALUE 20 mA (4003) = not equal to zero flow (e.g.
Page 41
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → CURRENT OUTPUT 1 → CONFIGURATION (only with PROFIBUS DP) VALUE 20 mA Use this function to assign the 20 mA current a value. (4003) The value can be greater or less than the value assigned to 0/4 mA, (function VALUE 0_4 mA (4002) →...
Page 42
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → CURRENT OUTPUT 1 → CONFIGURATION (only with PROFIBUS DP) MEASURING MODE Use this function to define the measuring mode for the current output. (4004) Options:...
Page 43
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → CURRENT OUTPUT 1 → CONFIGURATION (only with PROFIBUS DP) MEASURING MODE • PULSATING FLOW (Continued) If flow is characterized by severe fluctuations as is the case, for example, with reciprocating pumps, flow components outside the measuring range are buffered, balanced and output after a maximum delay of 60 seconds.
Page 44
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → CURRENT OUTPUT 1 → CONFIGURATION (only with PROFIBUS DP) Detailed explanations and 2. Defined measuring range ( – have different signs information I [mA] (Continued) A0001272 Flow a (—) outside, b (- -) within the measuring range.
Page 45
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → CURRENT OUTPUT 1 → CONFIGURATION (only with PROFIBUS DP) FAILSAFE MODE For safety reasons it is advisable to ensure that the current output assumes a predefined (4006) state in the event of a fault.
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6.1.2 Function group OPERATION → → OUTPUTS CURRENT OUTPUT 1 CONFIGURATION ↓ OPERATION Functional description OUTPUTS → CURRENT OUTPUT 1 → OPERATION (only with PROFIBUS DP) ACTUAL CURRENT Use this function to view the computed actual value of the output current.
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Group PULSE/FREQUENCY OUTPUT 1 6.2.1 Function group CONFIGURATION → OUTPUTS CURRENT OUTPUT 1 ↓ → PULSE/FREQ. 1 to 2 CONFIGURATION Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (GENERAL/ FREQUENCY)
Page 49
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY) (only with PROFIBUS DP) END VALUE FREQ. Note! (4203) This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function (4200).
Page 50
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY) (only with PROFIBUS DP) VALUE-f HIGH Note! (4205) This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function (4200).
Page 51
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY) (only with PROFIBUS DP) VALUE-f HIGH Parameter setting example 1: (Continued) 1. VALUE-f LOW (4204) = not equal to zero flow (e.g. –5 m VALUE-f HIGH (4205) = not equal to zero flow (e.g.
Page 52
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY) (only with PROFIBUS DP) MEASURING MODE Note! (4206) This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function (4200).
Page 53
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY) (only with PROFIBUS DP) MEASURING MODE • PULSATING FLOW (Continued) If flow is characterized by severe fluctuations as is the case, for example, with reciprocating pumps, flow components outside the measuring range are buffered, balanced and output after a maximum delay of 60 seconds.
Page 54
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY) (only with PROFIBUS DP) OUTPUT SIGNAL Note! (4207) Function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function (4200).
Page 55
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY) (only with PROFIBUS DP) OUTPUT SIGNAL Example for output configuration PASSIVE-POSITIVE: (Continued) Output configuration with an external pull-down resistance.
Page 56
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY) (only with PROFIBUS DP) OUTPUT SIGNAL Example for active output circuit (ACTIVE): (Continued) With an active circuit, the internal power supply is 24 V.
Page 57
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY) (only with PROFIBUS DP) TIME CONSTANT Note! (4208) This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function (4200).
Page 58
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (PULSE) (only with PROFIBUS DP) ASSIGN PULSE Note! (4221) This function is not available unless the PULSE setting was selected in the OPERATION MODE function (4200).
Page 59
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (PULSE) (only with PROFIBUS DP) PULSE WIDTH Note! (4223) This function is not available unless the PULSE setting was selected in the OPERATION MODE function (4200).
Page 60
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (PULSE) (only with PROFIBUS DP) MEASURING MODE Note! (4225) This function is not available unless the PULSE setting was selected in the OPERATION MODE function (4200).
Page 61
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (PULSE) (only with PROFIBUS DP) OUTPUT SIGNAL Note! (4226) Function is not available unless the PULSE setting was selected in the OPERATION MODE function (4200).
Page 62
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (PULSE) (only with PROFIBUS DP) OUTPUT SIGNAL Example for output configuration PASSIVE-POSITIVE: (Continued) Output configuration with an external pull-down resistance.
Page 63
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (PULSE) (only with PROFIBUS DP) OUTPUT SIGNAL Example for active output circuit (ACTIVE): (Continued) With an active circuit, the internal power supply is 24 V.
Page 64
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (PULSE) (only with PROFIBUS DP) FAILSAFE MODE Note! (4227) This function is not available unless the PULSE setting was selected in the OPERATION MODE function (4200).
Page 65
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS→ PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (STATUS) (only with PROFIBUS DP) ASSIGN STATUS Note! (4241) This function is not available unless the STATUS setting was selected in the OPERATION MODE function (4200).
Page 66
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS→ PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (STATUS) (only with PROFIBUS DP) SWITCH-ON DELAY Note! (4243) This function is not available unless STATUS was selected in the OPERATION MODE function (4200) and LIMIT VALUE or FLOW DIRECTION was selected in the ASSIGN STATUS function (4241).
Page 67
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS→ PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (STATUS) (only with PROFIBUS DP) MEASURING MODE Note! (4246) This function is not available unless the STATUS setting was selected in the OPERATION MODE function (4200) and the status output was assigned a limit value.
Page 68
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS→ PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (STATUS) (only with PROFIBUS DP) TIME CONSTANT Note! (4247) This function is not available unless the STATUS setting was selected in the OPERATION MODE function (4200).
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS 6.2.2 Function group OPERATION → OUTPUTS CURRENT OUTPUT 1 ↓ → PULSE/FREQ. 1 to 2 CONFIGURATION ↓ OPERATION Functional description OUTPUTS→ PULSE/FREQUENCY OUTPUT 1 → OPERATION (FREQUENCY) (only with PROFIBUS DP)
Page 70
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS→ PULSE/FREQUENCY OUTPUT 1 → OPERATION (PULSE) (only with PROFIBUS DP) SIMULATION PULSE Note! (4322) This function is not available unless the PULSE setting was selected in the OPERATION MODE function.
Page 71
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS→ PULSE/FREQUENCY OUTPUT 1 → OPERATION (PULSE) (only with PROFIBUS DP) ACTUAL STATUS Note! (4341) This function is not available unless the STATUS setting was selected in the OPERATION MODE function (4200).
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Group RELAY OUTPUT (1 to 2) 6.3.1 Function group CONFIGURATION → OUTPUTS CURRENT OUTPUT 1 ↓ PULSE/FREQ. 1 to 2 ↓ → RELAY OUTPUT 1 CONFIGURATION ↓ → RELAY OUTPUT 2...
Page 74
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → RELAY OUTPUT (1 to 2) → CONFIGURATION (only with PROFIBUS DP) ON-VALUE Note! (4701) This function is not available unless LIMIT or FLOW DIRECTION was selected in the ASSIGN RELAY function (4700).
Page 75
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS Functional description OUTPUTS → RELAY OUTPUT (1 to 2) → CONFIGURATION (only with PROFIBUS DP) SWITCH-OFF DELAY Note! (4704) This function is not available unless LIMIT was selected in the ASSIGN RELAY function (4700).
Page 76
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → RELAY OUTPUT (1 to 2) → CONFIGURATION (only with PROFIBUS DP) TIME CONSTANT Use this function to enter a time constant defining how the measuring signal reacts to...
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS 6.3.2 Function group OPERATION → OUTPUTS CURRENT OUTPUT 1 ↓ PULSE/FREQ. 1 to 2 ↓ → RELAY OUTPUT 1 CONFIGURATION ↓ ↓ OPERATION → RELAY OUTPUT 2 CONFIGURATION ↓...
Page 78
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description OUTPUTS → RELAY OUTPUT (1 to 2) → OPERATION (only with PROFIBUS DP) VALUE SIM. SWITCH Note! POINT The function is not visible unless the SIMULATION SWITCH POINT function (4741) is (4742) active (= ON).
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6.3.4 Response of the relay output General If you have configured the relay output signal for “LIMIT VALUE” or “FLOW DIRECTION”, you can define the requisite switch points in the ON-VALUE and OFF-VALUE functions. When the measured variable in question reaches one of these predefined values, the relay output switches as shown in the illustrations below.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 6 Block OUTPUTS 6.3.5 Switching action of the relay output Contact* Function Status Relay coil ON (operation) System in measuring mode XXX.XXX.XX energized A0001052 A0001239 A0001237 System not in measuring mode (power supply failed) XXX.XXX.XX...
Page 82
6 Block OUTPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Contact* Function Status Relay coil Flow forward direction CH1, CH2, AVG. energized A0001241 A0001239 A0001237 reverse de-energized A0001242 A0001240 A0001238 Limit value Limit value not overshot or – Volume flow undershot –...
7 Block INPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Group STATUS INPUT 7.1.1 Function group CONFIGURATION → → INPUTS STATUS INPUT CONFIGURATION Functional description INPUTS → STATUS INPUT → CONFIGURATION (only with PROFIBUS DP) ASSIGN STATUS INPUT Use this function to assign a switching function to the status input.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 7 Block INPUTS 7.1.2 Function group OPERATION → → INPUTS STATUS INPUT CONFIGURATION ↓ OPERATION Functional description INPUTS → STATUS INPUT → OPERATION (only with PROFIBUS DP) ACTUAL Use this function to view the current level of the status input.
7 Block INPUTS Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 7.1.3 Function group INFORMATION → → INPUTS STATUS INPUT CONFIGURATION ↓ OPERATION ↓ INFORMATION Functional description INPUTS → STATUS INPUT → INFORMATION (only with PROFIBUS DP) TERMINAL NO. Use this function to display the numbers of the terminals (in the connection (5080) compartment) which are used by the status input.
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Group PROFIBUS DP / PROFIBUS PA 8.1.1 Function group CONFIGURATION PROFIBUS-DP → → BASIC FUNCTION (depends on PROFIBUS module) CONFIGURATION PROFIBUS-PA Functional description BASIC FUNCTION → PROFIBUS-DP / PROFIBUS-PA → CONFIGURATION TAG NAME Use this function to assign a tag name to the measuring device.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8 Block BASIC FUNCTION 8.1.2 Function group FUNCTION BLOCKS PROFIBUS-DP → → BASIC FUNCTION (depends on PROFIBUS module) CONFIGURATION ↓ PROFIBUS-PA FUNCTION BLOCKS Functional description BASIC FUNCTION → PROFIBUS-DP / PROFIBUS-PA → FUNCTION BLOCKS...
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8.1.3 Function group TOTALIZER PROFIBUS-DP → → BASIC FUNCTION (depends on PROFIBUS module) CONFIGURATION ↓ PROFIBUS-PA FUNCTION BLOCKS ↓ TOTALIZER Functional description BASIC FUNCTION → PROFIBUS-DP / PROFIBUS-PA → TOTALIZER SELECT TOTALIZER Use this function to select a totalizer.
Page 91
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8 Block BASIC FUNCTION Functional description BASIC FUNCTION → PROFIBUS-DP / PROFIBUS-PA → TOTALIZER UNIT TOTALIZER Use this function to define a unit for the measured variable of the totalizer. (6134) Options: Metric →...
Page 92
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description BASIC FUNCTION → PROFIBUS-DP / PROFIBUS-PA → TOTALIZER TOTALIZER MODE Use this function to define how the totalizer totals the flow components. (6137) Options: BALANCE Positive and negative flow components. The positive and negative flow components are balanced.
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8.1.5 Function group INFORMATION PROFIBUS-DP → → BASIC FUNCTION (depends on PROFIBUS module) CONFIGURATION ↓ PROFIBUS-PA FUNCTION BLOCKS ↓ TOTALIZER ↓ OPERATION ↓ INFORMATION Functional description BASIC FUNCTION → PROFIBUS-DP / PROFIBUS-PA → FUNCTION BLOCKS PROFILE VERSION Use this function to view the profile version.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8 Block BASIC FUNCTION Group PROCESS PARAMETER (CH1 to CH2) 8.2.1 Function group CONFIGURATION PROFIBUS-DP → BASIC FUNCTION (depends on PROFIBUS module) PROFIBUS-PA ↓ PROCESS PARA. (CH1 to → CONFIGURATION CH2) Functional description BASIC FUNCTION →...
Page 96
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description BASIC FUNCTION → PROCESS PARAMETER (CH1 to CH2) → CONFIGURATION PRESSURE SHOCK The closure of a valve can cause brief but severe movements of the fluid in the piping SUPPRESSION system, movements which the measuring system registers.
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8.2.3 Function group LIQUID DATA PROFIBUS-DP → BASIC FUNCTION (depends on PROFIBUS module) PROFIBUS-PA ↓ PROCESS PARAMETER → CONFIGURATION ↓ ADJUSTMENT ↓ ↓ LIQUID DATA PROCESS PARAMETER →...
Page 99
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8 Block BASIC FUNCTION Functional description BASIC FUNCTION → PROCESS PARAMETER (CH1 to CH2) → LIQUID DATA SOUND VELOCITY This function displays the sound velocity of the liquid. This is determined via the values LIQUID entered in the LIQUID (6540) and TEMPERATURE (6541) functions.
Page 100
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description BASIC FUNCTION → PROCESS PARAMETER (CH1 to CH2) → LIQUID DATA SOUND VELOCITY Use this function to specify the upper search range for the sound velocity of the liquid.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8 Block BASIC FUNCTION Group SYSTEM PARAMETER (CH1 to CH2) 8.3.1 Function group CONFIGURATION PROFIBUS-DP → BASIC FUNCTION (depends on PROFIBUS module) PROFIBUS-PA ↓ PROCESS PARAMETER ↓ PROCESS PARAMETER ↓ SYSTEM PARAMETER →...
Page 102
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description BASIC FUNCTION → SYSTEM PARAMETER (CH1…CH2) → CONFIGURATION POSITIVE ZERO RETURN Use this function to interrupt evaluation of measured variables. This is necessary when a (6605) piping system is being cleaned, for example.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8 Block BASIC FUNCTION Group SENSOR DATA (CH1 to CH2) 8.4.1 Function group CONFIGURATION PROFIBUS DP → BASIC FUNCTION (depends on PROFIBUS module) PROFIBUS PA ↓ PROCESS PARAMETER ↓ PROCESS PARAMETER ↓...
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8.4.2 Function group MEASURING TUBE PROFIBUS-DP → BASIC FUNCTION (depends on PROFIBUS module) PROFIBUS-PA ↓ PROCESS PARAMETER ↓ PROCESS PARAMETER ↓ SYSTEM PARAMETER ↓ SYSTEM PARAMETER ↓ →...
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8 Block BASIC FUNCTION 8.4.3 Function group SENSOR PARAMETER PROFIBUS-DP → BASIC FUNCTION (depends on PROFIBUS module) PROFIBUS-PA ↓ PROCESS PARAMETER ↓ PROCESS PARAMETER ↓ SYSTEM PARAMETER ↓ SYSTEM PARAMETER ↓ →...
Page 106
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description BASIC FUNCTION → SENSOR DATA (CH1 to CH2) → CONFIGURATION CABLE LENGTH Use this function to select the length of the sensor cable. (6883) Options: LENGTH 5m/15 feet...
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8 Block BASIC FUNCTION 8.4.4 Function group CALIBRATION DATA PROFIBUS-DP → BASIC FUNCTION (depends on PROFIBUS module) PROFIBUS-PA ↓ PROCESS PARAM. CH1 ↓ PROCESS PARAM. CH2 ↓ SYSTEM PARAM. CH1 ↓ SYSTEM PARAM. CH2 ↓...
8 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 8.4.5 Function group ORIG. FACT. CALIBR. PROFIBUS-DP → BASIC FUNCTION (depends on PROFIBUS module) PROFIBUS-PA ↓ PROCESS PARAMETER ↓ PROCESS PARAMETER ↓ SYSTEM PARAMETER ↓ SYSTEM PARAMETER ↓...
9 Block SUPERVISION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Group SYSTEM (SYSTEM CH2) 9.1.1 Function group CONFIGURATION → → SUPERVISION SYSTEM CONFIGURATION Functional description SUPERVISION → SYSTEM → CONFIGURATION ALARM DELAY Use this function to specify a time period for suppressing the appearance of fault or notice (8005) messages.
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 9 Block SUPERVISION 9.1.2 Function group OPERATION → → SUPERVISION SYSTEM CONFIGURATION ↓ ↓ OPERATION → SYSTEM CH2 OPERATION Functional description SUPERVISION → SYSTEM [CH2] → OPERATION ACTUAL SYSTEM Use this function to check the present system condition.
Page 112
9 Block SUPERVISION Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functional description SUPERVISION → SYSTEM [CH2] → OPERATION VALUE SIMULATION Note! MEASURAND The function is not visible unless the function SIMULATION MEASURAND (8043) is (8044) active. Use this function to specify a selectable value (e.g. 12 m /s).
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 9 Block SUPERVISION Group VERSION INFO 9.2.1 Function group DEVICE → SUPERVISION SYSTEM ↓ → VERSION INFO DEVICE Functional description SUPERVISION → VERSION-INFO → DEVICE DEVICE SOFTWARE Displays the current device software version.
• The language options of the available language group are displayed in the LANGUAGE function (2000). • You can change the language group via the configuration software FieldCare. Please do not hesitate to contact your Endress+Hauser sales office if you have any questions. 9.2.4 Function group F-CHIP →...
10 Factory settings Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 10.1.3 Length, temperature Unit Length Temperature °C 10.2 US units (for USA and Canada only) 10.2.1 Low flow cutoff, totalizer Nominal diameter Low flow cut off Totalizer (approx. v = 0.04 m/s)
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 11 Index function matrix Index function matrix Blocks Function groups A = MEASURED VARIABLES ..... . 10 000 = MAIN VALUES .
Page 118
11 Index function matrix Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Functions 0... 4040 = ACTUAL CURRENT ..... . .46 4041 = SIMULATION CURRENT .
Page 119
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA 11 Index function matrix Functions 6... Functions 8... 6100 = TAG NAME ......88 8005 = ALARM DELAY .
Page 120
11 Index function matrix Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA Endress+Hauser...