8 Block BASIC FUNCTION
108
8.4.5
Function group ORIG. FACT. CALIBR.
→
BASIC FUNCTION
G
BASIC FUNCTION → SENSOR DATA (CH1 to CH2) → ORIG. FACTORY CALIB.
CALIBRATION DATE
This function resets the calibration data of the measuring device to the factory settings.
(6910)
Procedure:
1. Enter current date.
2. Store it.
The measuring device resets the calibration data to the factory settings and restarts
automatically.
!
• The calibration data reset is recorded in the calibration history.
• The date in the CALIBRATION DATE function (6808) is overwritten.
User input:
Format depends on the selection in function FORMAT DATE/TIME (0429)
Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA