Technical Data
72
10.6

Mechanical construction

Design, dimensions
Weight
Transmitter materials
10.7

Measurement data of CLS 52 sensor

Conductivity measuring
range
Measured value deviation
Cell constant
Temperature sensor
Temperature measuring
range
Temperature response time Temperature response time t
Materials in contact with
medium
10.8

Process

Process temperature
Sterilisation
Process pressure
Ingress protection CLS 52
sensor
Separate transmitter with
mounting plate:
Compact transmitter MV1, CS1,
GE1, SMS versions:
Compact transmitter VA1, AP1
versions:
Separate version:
Compact version with CLS 52
sensor:
Housing
Front window
10 µS/cm ... 2000 mS/cm
-5 ... 100 °C / 23 ... 212 °F
> 100 °C / 212 °F
–1
k = 5.9 cm
Pt 100 (class A acc. to IEC 60751)
–35 ... +140 °C / -31 ... +284 °F
90
PEEK
stainless steel 1.4435 (AISI 316L)
®
Chemraz
CLS 52 sensor with separate
version
Compact version
CLS 52 sensor with separate
version
Compact version
max. 16 bar (90 °C) / 232 psi (194 °F)
IP 67 / NEMA 6 (fixed cable)
SmarTec S CLD 132
L x W x D: 225 x 142 x 109 mm /
8.86 x 5.59 x 4.29"
L x W x D: 225 x 142 x 242 mm /
8.86 x 5.59 x 9.53"
L x W x D: 225 x 142 x 180 mm /
8.86 x 5.59 x 7.09"
approx. 2.5 kg / 5.5 lb.
approx. 3 kg / 6.6 lb.
stainless steel 1.4301, polished
polycarbonate
±10 µS/cm + 0.5 % of measured value
±30 µS/cm + 0.5 % of measured value
< 5 s
max. 125 °C / 257 °F at 70 °C / 158 °F
ambient temperature
max. 55 °C / 131 °F at 55 °C / 131 °F
ambient temperature
140 °C / 284 °F at 70 °C / 158 °F ambient
temperature, 4 bar / 58 psi, max. 30 min
140 °C / 284 °F at 35 °C / 95 °F ambient
temperature, max. 30 min