Device Functions Proline Promass 83 FOUNDATION Fieldbus
Endress+Hauser
7.4

Group SENSOR DATA

7.4.1

Function group CONFIGURATION

BASIC FUNCTION
G
BASIC FUNCTION → SENSOR DATA → CONFIGURATION
All sensor data (calibration factor, zero point and nominal diameter) are set at the factory and saved on the S-DAT
sensor memory chip.
"
Caution!
Under normal circumstances you should not change the following parameter settings, because changes affect numerous
functions of the entire measuring facility in general and the accuracy of the measuring system in particular. For this
reason, the functions described below cannot be changed even when you enter your personal code.
Contact the Endress+Hauser service organization if you have any questions about these functions.
!
Note!
The individual values of the functions are also provided on the sensor nameplate.
K-FACTOR
(6800)
ZERO POINT
(6803)
NOMINAL DIAMETER
(6804)
FOUNDATION FIELDBUS GGA
PROCESS PARAMETER
SYSTEM PARAMETER
SENSOR DATA
Function description
This function shows the current calibration factor for the sensor.
Factory setting:
Depends on nominal diameter and calibration.
This function shows the current zero-point correction value for the sensor.
User input:
Max. 5-digit number: –99999 to +99999
Factory setting:
Depends on calibration
This function shows the nominal diameter for the sensor.
Factory setting:
Depends on nominal diameter
7 Block BASIC FUNCTION
GIA
GLA
GNA
CONFIGURATION
680
61