Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA
Endress+Hauser
8.2.2

Function group ADJUSTMENT

BASIC FUNCTION
G
BASIC FUNCTION → PROCESS PARAMETER (CH1 to CH2) → ADJUSTMENT
ZERO POINT
This function enables a zero point adjustment to be automatically carried out. The new

ADJUSTMENT

zero point determined by the measuring system is adopted by the ZERO POINT
function (6891) ( → Page 107).
(6480)
User input:
CANCEL
START
Factory setting:
CANCEL
"
Before carrying this out, please refer to the Operating Instructions for
Prosonic Flow 93C PROFIBUS DP/PA, BA 089D, for a detailed description of the
procedure for zero point adjustment.
!
• Programming is locked during zero point adjustment. The message "ZERO ADJUST
• If the zero point adjustment is not possible, e.g. with a flow velocity > 0.1 m/s, or has
PROFIBUS-DP
(depends on PROFIBUS module)
PROFIBUS-PA
PROCESS PARAMETER
CH1
PROCESS PARAMETER
CH2
Functional description
Caution!
Note!
RUNNING" appears on the display.
been canceled, then the alarm message "ZERO ADJUST NOT POSSIBLE" is shown on
the display.
8 Block BASIC FUNCTION
GBA
GCA
GIA
CONFIGURATION
ADJUSTMENT
GIB
CONFIGURATION
ADJUSTMENT
640
648
640
648
97

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline Prosonic Flow 93C and is the answer not in the manual?

Questions and answers

Table of Contents