Device functions Proline Prosonic Flow 93C PROFIBUS DP/PA
Endress+Hauser
8.2.2
Function group ADJUSTMENT
→
BASIC FUNCTION
G
BASIC FUNCTION → PROCESS PARAMETER (CH1 to CH2) → ADJUSTMENT
ZERO POINT
This function enables a zero point adjustment to be automatically carried out. The new
ADJUSTMENT
zero point determined by the measuring system is adopted by the ZERO POINT
function (6891) ( → Page 107).
(6480)
User input:
CANCEL
START
Factory setting:
CANCEL
"
Before carrying this out, please refer to the Operating Instructions for
Prosonic Flow 93C PROFIBUS DP/PA, BA 089D, for a detailed description of the
procedure for zero point adjustment.
!
• Programming is locked during zero point adjustment. The message "ZERO ADJUST
• If the zero point adjustment is not possible, e.g. with a flow velocity > 0.1 m/s, or has
PROFIBUS-DP
(depends on PROFIBUS module)
PROFIBUS-PA
↓
PROCESS PARAMETER
CH1
↓
PROCESS PARAMETER
CH2
Functional description
Caution!
Note!
RUNNING" appears on the display.
been canceled, then the alarm message "ZERO ADJUST NOT POSSIBLE" is shown on
the display.
8 Block BASIC FUNCTION
GBA
GCA
→
GIA
CONFIGURATION
↓
ADJUSTMENT
→
GIB
CONFIGURATION
↓
ADJUSTMENT
640
648
640
648
97