Configuration using Rosemount TankMaster

The Rosemount TankMaster Winsetup program is the recommended configuration tool
for the Rosemount 5900S. Typically a Rosemount 2410 Tank Hub is connected to a
Rosemount 2460 System Hub which communicates with host system via TRL2 Modbus,
RS485 Modbus, Modbus TCP, or emulation protocol. A Rosemount 5900S can be Installed
and configured by one of the following methods:
as part of the installation and configuration procedure of a Rosemount 2410 Tank Hub
(recommended)
by using the Rosemount TankMaster installation wizard
A Rosemount 5900S level gauge is typically installed as part of the installation procedure
when installing a Rosemount 2410 Tank Hub in Rosemount TankMaster WinSetup. Then
the level gauge appears in the WinSetup workspace and is configured in a separate stage
via the Properties window.
See the Rosemount Tank Gauging
configure a Rosemount 5900S Radar Level Gauge.
4.3.1
Installation wizard
The Rosemount TankMaster WinSetup installation wizard is a tool that facilitates
installation and configuration of Rosemount 5900S and other devices. This may be useful
in case the Rosemount 5900S was not installed as part of the installation procedure for the
Rosemount 2410.
See the Rosemount Tank Gauging
Note
In case the Rosemount 5900S level gauge was installed "offline" via a Rosemount 2410
Tank Hub, it needs to be configured separately via the Properties window.
To install a Rosemount 5900S by using the Rosemount TankMaster WinSetup wizard do
the following:
Procedure
1. Start the TankMaster WinSetup program.
2. Select the Devices folder.
3. Click the right mouse button and select Install New.
4. Follow the instructions.
Rosemount 5900S Radar Level Gauge
There are a number of configuration options available which are not included in the
installation wizard. See
Basic configuration
information on how to use various options such as Tank Scan, Empty Tank
Handling, Surface Echo Tracking and Filter Settings.
Configuration Manual
for more information on how to
System Configuration
Manualfor more information.