Echo curve
The Echo Curve window lets you analyze the measurement signal from a Rosemount
5900S. It allows you to view tank echoes and configure parameters to enable the gauge to
distinguish between surface echoes and disturbing echoes and noise. For details see
Scan.
Figure 4-22: Echo Curve Configuration
The Echo Peaks button lets you open the Echo Peaks window which allows you to register
false echoes.
The Echo Threshold Settings button opens the Echo Threshold Settings window which lets
you set a general amplitude threshold to filter out noise. You may also create a
customized amplitude threshold curve to optimize disturbing echo filtering.
For more information see chapter "Service Functions/Tank Scan" in the Rosemount Tank
Gauging
Rosemount 5900S Radar Level Gauge
System Configuration
Manual.
Configuration
September 2019