Emerson Rosemount 5900S Reference Manual page 211

Radar level gauge
Hide thumbs Also See for Rosemount 5900S:
Table of Contents

Advertisement

Reference Manual
00809-0100-5900
Table 6-4: Error Messages for the Rosemount 5900S
Message
Database (Hreg) error
Other Memory error
Microwave Module error
RM error
Other hardware error
Rosemount 5900S Radar Level Gauge
(continued)
Description
Input register no. 6004.
An error in the transmitter
configuration memory (EEPROM) has
been detected. The error is either a
checksum error that can be solved by
loading the default database or a
hardware error.
Note
The default values are used until the
problem is solved.
The following bits indicate a serious
Holding register problem:
Bit 0: DSP checksum
Bit 1: DSP limit
Bit 2: DSP version
Bit 3: Write error
Input register no. 6006.
Input register no. 6008.
Bit 0: Not connected
Input register no. 6010
Bit 1: SW configuration
Bit 5: FPROM Checksum
Bit 6: FPROM Version
Bit 9: HREG Checksum
Bit 10: HREG Limit
Bit 11: HREG Write
Bit 12: HREG Read
Bit 13: HREG Version
Bit 14: MWM Invalid Id
Bit 30: SW Serious Error
Input register no. 6022.
An unspecified hardware error has
been detected.
Bit 0: Internal Temp Out of Range
Service and troubleshooting
September 2019
Action
Load default database and restart the
Rosemount 5900S Radar Level Gauge.
Contact Emerson Automation
Solutions /Rosemount Tank Gauging
service department if the problem
persists.
Contact Emerson Automation
Solutions /Rosemount Tank Gauging
service department.