Service and troubleshooting
September 2019
6.2

Service

This section briefly describes functions which may be useful for service and maintenance
of a Rosemount 5900S Radar Level Gauge. If not otherwise stated, most examples are
based on using the Rosemount TankMaster WinSetup tool to access these functions. See
the Rosemount Tank Gauging
to use the WinSetup program.
6.2.1
Viewing input and holding registers using TankMaster
In a Rosemount Tank Gauging system, measurement data is continuously stored in Input
Registers of devices such as the Rosemount 2410 Tank Hub, Rosemount 5900 Radar Level
Gauge, and other devices. By viewing the input registers of a device, you can verify that
the device is working properly.
Holding Registers store various device parameters used to control measurement
performance.
Procedure
1. Start the TankMaster WinSetup program.
2. In the TankMaster WinSetup workspace window, select the device icon.
3. Right-click and select View Input/View Holding Registers option, or from the
188
System Configuration Manual
Service menu choose Devices → View Input/View Holding Registers.