Brazil; September - Emerson Rosemount 5900S Reference Manual

Radar level gauge
Hide thumbs Also See for Rosemount 5900S:
Table of Contents

Advertisement

Product Certifications

September 2019

2. Non-metallic surfaces and the surface of the painted housing may, under certain
3. Using the box provided on the nameplate, the User shall permanently mark the
4. When installed as Ex ib Ga/Gb, the partition wall materials separating EPL Ga from
5. Maximum Process Temperatures are as follows:
B.9

Brazil

B.9.1
I2 INMETRO Intrinsic Safety
Certificate
Standards
Markings
Entity parameters
FISCO parameters
Special Conditions for Safe Use (X):
1. See certificate for special conditions.
258
extreme conditions, generate an ignition-capable level of electrostatic. Appropriate
measures must be taken to prevent electrostatic discharge.
type of protection chosen for the specific installation. Once the type of protection
has been marked it shall not be changed.
EPL Gb are constructed of different materials depending on the antenna option.
Please refer to Control Drawing D9240040-917 for the material type of each
antenna. The material shall not be subject to environmental conditions which might
adversely affect the partition wall.
When option n=Tank Seal
PV or QV
PK, FK, HK or QK
PE or QE
PB or QB
PM, FF, HH or QM
PF or QF
UL-BR 17.0982X
ABNT NBR IEC 60079-0:2013, 60079-11:2013, 60079-26:2016
Ex ia IIC T4 Ga/Gb
Tamb: -50 °C to + 80 °C
IP66/IP67
Ui (Vmax)
30 V
17.5V
O-ring Type
Viton
Kalrez
EPDM
BUNA-N
FVMQ
FEP
Ii (lmax)
Pi
300 mA
1.3 W
380 mA
5.32 W
Reference Manual
00809-0100-5900
Min/Max Process
Temperature Range
-15°C to +180°C
-20°C to +230°C
-40°C to +110°C
-35°C to +90°C
-60°C to +155°C
-60°C to +180°C
Ci
Li
1.1 nF
1.5 µH
1.1 nF
1.5 µH
Reference Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents