Advertisement

Reference Manual
00809-0100-5900
5. Enter hand dipped level values and the corresponding levels measured by the
6. Click the Refresh button. Now WinSetup calculates the deviations between hand
7. Click the Save Calibration Data in PC Database button in order to save the entered
Rosemount 5900S Radar Level Gauge
Figure 4-12: Calibration Data Window
A. Level gauge
B. Hand dip
Rosemount 5900S level gauge. It is recommended that the hand dipped levels are
based on the average value of three consecutive measurements within 1 mm. For
further information see
Hand
Note
Measurement unit mm is used in the Calibration Data window.
dipped and measured levels.
values and return to the Calibrate window.
The Calibrate window displays a straight line fitted through measurement points
that represent the difference between hand dipped level values, and values
measured by the level gauge. For still pipe antennas a sloping line is displayed,
otherwise the line is horizontal. The slope is due to the linear impact by the still pipe
on the microwave velocity of propagation.
dipping.
Configuration
September 2019
131