Emerson Rosemount 5900S Reference Manual page 282

Radar level gauge
Hide thumbs Also See for Rosemount 5900S:
Table of Contents

Advertisement

F
Fieldbus Block Information
OUNDATION
September 2019
Table C-5: Measurement Transducer Block Parameters
Index
Number
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
Table C-6: Device Mode
VALUE
0
2
3
Table C-7: Antenna Type
VALUE
5001
5002
3002
2001
6001
282
Parameter
TANK_EXPANSION_COEFF
TANK_CALIB_AVG_TEMP
DAMPING_VALUE
HEART_BEAT_COUNT
DEVICE_STATUS
DEVICE_COMMAND
VOLUME
VOLUME_UNIT
MODEL_CODE
FF_SUPPORT_INFO
FF_APPL_VERSION_NUMBER CM version number
SENSOR_DIAGNOSTICS
VAPOR_PRESSURE
VAPOR_TEMPERATURE
USER_DEFINED
TANK_TEMPERATURE
PRESSURE_UNIT
USED_HOLD_OFF
ENV_DEVICE_MODE
Normal operation
Restart device
Set device to factory default
ANTENNA_TYPE
Still-Pipe Array Fixed
Still-Pipe Array Hatch
Parabolic
Horn
LPG/LNG 150 psi Valve
(continued)
Description
Tank Expansion Coefficient
Tank Calibration Average Temperature
Damping Value
This number should be incrementing. It is an indication
that the device is alive.
Device Status. See also
Device
Command
Product volume in the tank. A value of 0 could indicate
that the volume calculation is not enabled.
Unit code for all volume parameters
Model code
FF Support Information
Sensor Diagnostics
Tank vapor pressure. Data provided from AO block.
Tank vapor temperature. Data provided from AO block.
User Defined value
Tank Temperature
Pressure Unit
Used Hold Off Distance
Reference Manual
00809-0100-5900
status.
Reference Manual

Hide quick links:

Advertisement

Table of Contents
loading