Advertisement

Configuration
September 2019
Procedure
1. Record the hand dipped ullage values and the corresponding values measured by
2. Enter the hand dipped level values and the level gauge values into the WinSetup
3. Inspect the resulting calibration graph and, if necessary, exclude measurement
4.7.3
To enter calibration data
Prerequisites
Make sure that the following information is available when you intend to use the Calibrate
function in Rosemount TankMaster WinSetup:
A list of hand dipped ullage values.
A list of level values measured by the Rosemount 5900S that correspond to the hand
dipped ullage/level values.
Procedure
1. In the Rosemount TankMaster WinSetup workspace window select the Rosemount
2. Click the right mouse button and choose Calibrate, or choose Calibrate from the
3. The Calibrate window is empty before any data is entered. Ensure that the gauge
4. Click the Calibration Data button.
130
the level gauge.
Calibration Data window (see
points which should not be used in the adjustment calculation.
5900S level gauge to be calibrated.
Service/Devices menu.
communicates properly with TankMaster by verifying that the Tank Reference
Height appears in the lower left corner.
To enter calibration
data).