Reference Manual
00809-0100-5900
Configure the verification pin
Prerequisites
Make sure that the position of the Verification Pin is accurately measured and the inner
diameter of the still-pipe is available.
Note
When the product surface is close to a Verification Pin, the radar echoes from the
Verification Pin and the product surface interfere. This may reduce the accuracy of the
measured distance to the Verification Pin. It is recommended that verification is not
performed if the distance between the Verification Pin and the product surface is less than
900 mm (see
A. Verification Pin
B. Minimum clearance 900 mm
Since hand dipping can not be performed in high pressurized tanks, Emerson Automation
Solutions/ Rosemount Tank Gauging has developed a unique method to verify level
gauging in such tanks. The method is based on measurements in a special radar wave
propagation mode against a fixed Verification Pin in order to verify the measurement.
Procedure
1. In the Rosemount
2. Select the icon that represents the desired radar level gauge.
Rosemount 5900S Radar Level Gauge
LPG/LNG antenna
requirements).
TankMaster WinSetup work space, select the Logical View tab.
Configuration
September 2019
123