Communication / Display / Configuration; Foundation ™ Fieldbus Characteristics - Emerson Rosemount 5900S Reference Manual

Radar level gauge
Hide thumbs Also See for Rosemount 5900S:
Table of Contents

Advertisement

Specifications and reference data
September 2019
A.1.8
Installation considerations
See
Installation
A.1.9
Measurement principle
The FMCW-method (Frequency Modulated Continuous Wave) means that the transmitted
radar signal has a linear frequency variation around 10 GHz. The reflection from the liquid
surface has a slightly different frequency compared with the signal transmitted from the
antenna when the reflection is received. The difference in frequency is directly
proportional to the distance between the antenna and the liquid surface, and thereby also
the liquid level. This technology enables a very accurate and stable measured value.
Figure A-1: Principle of FMCW Technology
A. Frequency, f (GHz)
B. Time, t (s)
A.2

Communication / Display / Configuration

A.2.1
Output variables and units
Level, and ullage: meter, centimeter, millimeter, feet, or inch
Level rate: meter/second, meter/hour, feet/second, feet/hour, inch/minute
Signal strength: mV
A.2.2
Configuration tools
Rosemount TankMaster WinSetup, Field Communicator
A.3
F
OUNDATION
Polarity sensitive
No
Quiescent current draw
51 mA