Installation
September 2019
3.4.9
Rosemount 5900S in a F
The Rosemount 5900S Radar Level Gauge supports the F
technology and lets you integrate it into an existing FF network. As long as the power
supply meets certain requirements (see
5900S
Figure 3-21: Example of an I.S. F
A. I.S. Power Supply
B. Trunk
C. Rosemount 2230 Display
D. Segment Coupler
E. Rosemount 644 Temperature Transmitter
F. Rosemount 5900S Radar Level Gauge
Ensure that the power supply is able to provide the total current needed for all the
connected devices. See
Ensure that:
the Rosemount 5900S and other devices connected to the F
system are compliant with the FISCO or Entity parameters of the power supply.
the short circuit protection of the Segment Coupler
consumption of the connected devices.
(7)
See
Product Certifications
(8)
See the Rosemount 2410 Reference Manual (Document No. 300530EN) for more information on the Segment Coupler.
82
(7)
will be able to operate as any other FF device.
Power requirements
for Rosemount 5900S approval information
OUNDATION
OUNDATION
Figure 3-21
and
Figure
Fieldbus System
OUNDATION
and
Power budget
(8)
matches the current
Reference Manual
00809-0100-5900
Fieldbus system
Fieldbus (FF)
3-22) the Rosemount
for further information.
Fieldbus (FF)
OUNDATION
Reference Manual