Configuration
September 2019
Choose correction method
There are several options available depending on gas mixture in the tank.
Procedure
1. In the LPG Setup window click the Correction button to open the LPG Correction
2. Choose one of the following correction methods:
128
window:
Option
Description
Air Correction, LPG
This method should only be used when there is no vapor in
correction disabled
the tank, i.e. when the tank is empty and contains air only.
It is used in the initial step when calibrating the Rosemount
5900S.
One known gas
This method may be used when there is only one gas type
in the tank. It provides the highest accuracy among the
different correction methods. Note that even small
amounts of another gas reduces the accuracy.
One or more
Use this method for hydrocarbons, for example Propane/
unknown gases
Buthane, when the exact mixture is not known.
Two gases with
This method is suitable for a mixture of two gases even if
unknown mixratio
the mixratio is not known.
One or more known
This method may be used when there is a well known
gases with known
mixture of up to 4 products in the tank.
mixratio
Now the Rosemount 5900S level gauge is ready to measure the product level when