About Rosemount 470Xa Gas Chromatographs; Description; Specifications - Emerson Rosemount 470XA Manual

Gas chromatograph
Hide thumbs Also See for Rosemount 470XA:
Table of Contents

Advertisement

Manual
MS-00809-0100-0470
1 
About Rosemount 470XA Gas
Chromatographs
1.1 

Description

The Rosemount 470XA Gas Chromatograph (GC) is part of the XA series of Emerson gas
chromatographs. Designed to simplify natural gas measurement analysis, the Rosemount
470XA provides greater ease of use and increased measurement performance for your C6+
BTU/CV analysis.
Another unique benefit of the Rosemount 470XA is its Maintainable Module technology,
which allows you to easily replace the GC module in the field in approximately two hours,
including warm-up and purge, greatly reducing downtime and overall operating costs.
1.2 

Specifications

Table 1-1: Electronics specifications
Specification
Power supply
Power consumption at 72 °F (22 °C)
Table 1-2: Construction specifications
Specification
Environmental temperature
Enclosure protection rating
Dimensions
(without sample system or mounts)
Mounting options
Weight (without sample system or
mounts)
Table 1-3: Performance specifications
Specification
Applications
Rosemount 470XA
Description
24 Vdc (standard) at the unit
21-30 Vdc (operating range) at the unit
Class 2 and SELV as specified by CEC, C22.1, and NEC, National Fire Protection
Association (NFPA)
NOTICE
Provide the gas chromatograph (GC) with one 5-amp circuit breaker for
protection.
50 Watts (start-up)
20 Watts (steady state)
Description
-4 to +140 °F (-20 to +60 °C)
IP65 and Type 4X
18 in. (height) x 15.6 in. (width) x 11 in. (depth)
460 mm (height) x 396 mm (width) x 280 mm (depth)
Pipe, wall, or bench
50 lb. (23 kg)
Description
4 minute C6+ standard analysis

About Rosemount 470XA Gas Chromatographs

(1)
June 2023
9