Configuration
September 2019
8. Check that the line fits well to the measurement points. If a point deviates
9. Click the Write new calibration data to RTG button to save the current calibration
132
significantly from the line, it can be excluded from the calculations. Open the
Calibration Data window (click the Calibration Data button) and uncheck the
corresponding check box in the Enable column.
data to the level gauge database registers.
Note
By clicking the Write new calibration data to RTG button, the Level values in the
Calibration Data window are recalculated and the old Calibration Data is replaced.
Now you can check the calibration result in the Calibrate window again:
Note that all measured values are adjusted according to the calculated Calibration
Distance and Correction Factor. In the Calibration Data window you can also see that
the level values measured by the Rosemount 5900S gauge are adjusted. Of course,
the hand dipped levels are unaltered.
Note
Do not change the Calibration Distance in the Properties/Tank Geometry window
when calibration is finished.
Reference Manual
00809-0100-5900
Reference Manual