Figure 6-1: Log File
6.2.9
Loading the default database using TankMaster
The Default Database is the original factory settings of the holding register database.
Rosemount TankMaster WinSetup offers the option to load the Default Database. This can
be useful if, for example, you want to try new database settings and then want to reload
the original factory settings, or when tank conditions have been altered.
Prerequisites
If error messages appear or other problems occur concerning the Database a
troubleshooting is recommended before loading the Default Database.
Note
The device address remains unaltered when the default database is loaded.
Procedure
1. In the TankMaster WinSetup workspace window, select the desired device icon.
2. Right-click and select View Holding Register.