Saving The Instrument State; Recalling The Instrument State - Keysight V3500A User Manual

Handheld rf power meter
Hide thumbs Also See for V3500A:
Table of Contents

Advertisement

2
Knowing your Instrument

Saving the instrument state

N O T E
C A U T I O N

Recalling the instrument state

N O T E
20
• The default "Save State" menu setting is <No>.
Saving the state of the instrument places all menu values in the memory of the
instrument so that it can be recalled after the cycling power. Frequency settings
are not saved or recalled.
• The "Save State" feature is only functional when the batteries are stored in the
instrument.
1 Press MENU to enter menu mode.
2 Scroll using the down menu arrow keypad until "Save
State" menu is displayed.
All menu commands are executed immediately when the choice is
selected. Therefore, the Save State command is implemented as soon
as the <Save> menu item is highlighted. There is no undo option— as
soon as the <Save> menu item is selected, the current menu setting
overwrite the previously saved settings.
3 Use the right arrow menu keypad to select <Save>. This
will save the instrument state in the non- volatile memory
of the V3500A.
The default "Recall State" menu setting is <No>.
Recalling the instrument state restores all stored menu values in the
V3500A non-volatile memory. Frequency settings are not recalled.
1 Press MENU to enter menu mode.
2 Scroll using the down menu arrow keypad until "Recall
State" menu is displayed.
V3500A User's Guide