User Maintenance
Check Sheet
Check Sheet
Check Sheet Fill-Out Example
The following example shows how the check sheet is filled out following
electrode wear check. Fill out the sheet in the same manner for short plate
wear check.
Electrode Wear Check Fill-Out Example
Table 4-3
Reference Value and User Limit Values Fill-Out Example
Measurement
1
Frequency
Parameter
100 MHz
Rs
Ls
800 MHz
Rs
Ls
1. Set by the user as desired.
2.
Record values obtained at the time of
3.
"Example of User Limit Values Setting" on page 37
See also
Table 4-4
Check History Fill-Out Example
Date
Oct./11/1999
9:30
Oct./11/1999
9:35
Oct./12/1999
9:30
Oct./12/1999
9:35
42
2
Reference Value
[a]
90m
290 pH
310 m
260 pH
"Reference Value Acquisition" on page 39
Frequency
100 MHz
800 MHz
100 MHz
800 MHz
1. When the result is unacceptable, replace the part.
Keysight 16196A/B/C/D Parallel Electrode SMD Test Fixture
Lower Limit
3
User Limit Value
[a-b]
[b]
30m
60m
790pH
500pH
40m
270m
400pH
660pH
.
Measurement
Measured Value
parameter
Rs
100m
Ls
320pH
Rs
345m
360pH
Ls
Rs
105m
Ls
340pH
Rs
355m
320pH
Ls
Upper Limit
[a+b]
120m
210pH
350m
140pH
.
Pass/Fail
Pass
Pass
Pass
Pass
Pass
Pass
1
Fail
Pass