:POWer Commands
:POWer:HARMonics:APPLy
Command Syntax
:POWer:HARMonics:APPLy
The :POWer:HARMonics:APPLy command applies the current harmonics analysis.
Switching power supplies draw a range of harmonics from the AC mains.
Standard limits are set for these harmonics because these harmonics can travel
back to the supply grid and cause problems with other devices on the grid.
Use the Current Harmonics analysis to test a switching power supply's current
harmonics to pre-compliance standard of IEC61000-3-2 (Class A, B, C, or D). The
analysis presents up to 40 harmonics.
See Also
•
•
•
•
•
•
•
•
•
642
(see
page
1428)
":POWer:HARMonics:DATA"
":POWer:HARMonics:DISPlay"
":POWer:HARMonics:FAILcount"
":POWer:HARMonics:LINE"
":POWer:HARMonics:POWerfactor"
":POWer:HARMonics:STANdard"
":POWer:HARMonics:STATus"
":POWer:HARMonics:RUNCount"
":POWer:HARMonics:THD"
on page 643
on page 644
on page 645
on page 646
on page 647
on page 651
on page 652
on page 650
on page 653