:OPERegister:CONDition (Operation Status Condition Register)
Query Syntax
:OPERegister:CONDition?
The :OPERegister:CONDition? query returns the integer value contained in the
Operation Status Condition Register.
From Hardware
Event Registers
OR
To OPER bit in
Status Byte
Table 86 Operation Status Condition Register
Bit
14
13
12
11
10
9
8-6
5
4
3
2-0
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1428)
From Overload
From Mask Test
Event Registers
Event Registers
IOF
IOC
HWE
OVLR
MTE
14
13
12
11
IOF
IOC
HWE
OVLR
MTE
14
13
12
11
10
+
Register
Name
Description
IOF
IO Operation Failed
IOC
IO Operation Complete
HWE
Hardware Event
OVLR
Overload
---
---
MTE
Mask Test Event
---
---
Wait
Wait Trig
Trig
---
---
Run
Running
---
---
Arm
AER?
Reg
Wait
Run
Trig
9
5
3
Wait
Run
Trig
9
8
7
6
5
4
3
When Set (1 = High = True), Ind icates:
Event when the IO operation fails.
Event when the IO operation completes.
Event when hard ware event occurs.
A 50Ω input overload has occurred.
(Not used.)
A mask test event has occurred.
(Not used.)
The trigger is armed (set by the Trigger Armed
Event Register (TER)).
(Not used.)
The oscilloscope is running (not stopped).
(Not used.)
Root (:) Commands
Run bit set if oscilloscope not stopped
:OPEReration:CONDition?
Operation Status Condition Register
:OPERation[:EVENt]?
Operation Status Event Register
2
1
0
:OPEE
:OPEE?
Operation Status Enable
(Mask) Register
6
247
Need help?