:DIGital<d>:POSition
Command Syntax
:DIGital<d>:POSition <position>
<d> ::= 0 to (# digital channels - 1) in NR1 format
<position> ::= integer in NR1 format.
Channel Size
Large
Medium
Small
The :DIGital<d>:POSition command sets the position of the specified channel.
Note that bottom positions might not be valid depending on whether digital
buses, serial decode waveforms, or the zoomed time base are displayed.
This command is only valid for the MSO models.
NOTE
Query Syntax
:DIGital<d>:POSition?
The :DIGital<d>:POSition? query returns the position of the specified channel.
If the returned value is "-1", this indicates there is no space to display the digital
waveform (for example, when all serial lanes, digital buses, and the zoomed time
base are displayed).
Return Format
<position><NL>
<position> ::= integer in NR1 format.
See Also
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1428)
Position
0-7
0-15
0-31
"Introduction to :DIGital<d> Commands"
:DIGital<d> Commands
Top
Bottom
7
0
15
0
31
0
on page 352
14
355