Opee (Operation Status Enable Register) - Keysight Technologies InfiniiVision MSO-X 4022A Programmer's Manual

Hide thumbs Also See for InfiniiVision MSO-X 4022A:
Table of Contents

Advertisement

:OPEE (Operation Status Enable Register)

Command Syntax
:OPEE <mask>
<mask> ::= 15-bit integer
The :OPEE command sets a mask in the Operation Status Enable register. Set any
of the following bits to "1" to enable bit 7 in the Status Byte Register and
potentially cause an SRQ (Service Request) interrupt to be generated.
From Hardware
Event Registers
OR
To OPER bit in
Status Byte
Table 85 Operation Status Enable Register (OPEE)
Bit
14
13
12
11
10
9
8-6
5
4
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1428)
From Overload
From Mask Test
Event Registers
Event Registers
IOF
IOC
HWE
OVLR
MTE
14
13
12
11
OVLR
MTE
IOF
IOC
HWE
14
13
12
11
10
+
Register
Name
Description
IOF
IO Operation Failed
IOC
IO Operation Complete
HWE
Hardware Event
OVLR
Overload
---
---
MTE
Mask Test Event
---
---
Wait
Wait Trig
Trig
---
---
Arm
AER?
Reg
Wait
Run
Trig
9
5
3
Wait
Run
Trig
9
8
7
6
5
4
3
When Set (1 = High = True), Enables:
Event when the IO operation fails.
Event when the IO operation completes.
Event when hard ware event occurs.
Event when 50Ω input overload occurs.
(Not used.)
Event when mask test event occurs.
(Not used.)
Event when the trigger is armed.
(Not used.)
Root (:) Commands
Run bit set if oscilloscope not stopped
:OPEReration:CONDition?
Operation Status Condition Register
:OPERation[:EVENt]?
Operation Status Event Register
2
1
0
:OPEE
:OPEE?
Operation Status Enable
(Mask) Register
6
245

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the InfiniiVision MSO-X 4022A and is the answer not in the manual?

This manual is also suitable for:

Infiniivision 4000 x seriesInfiniivision mso-x 4032aInfiniivision mso-x 4034aInfiniivision mso-x 4054aInfiniivision mso-x 4104aInfiniivision mso-x 4154a ... Show all

Table of Contents