External:probe - Keysight Technologies InfiniiVision MSO-X 4022A Programmer's Manual

Hide thumbs Also See for InfiniiVision MSO-X 4022A:
Table of Contents

Advertisement

:EXTernal:PROBe

Command Syntax
:EXTernal:PROBe <attenuation>
<attenuation> ::= probe attenuation ratio in NR3 format
The :EXTernal:PROBe command specifies the probe attenuation factor for the
external trigger. The probe attenuation factor may be 0.1 to 1000. This command
does not change the actual input sensitivity of the oscilloscope. It changes the
reference constants for scaling the display factors and for setting trigger levels.
If an AutoProbe probe is connected to the oscilloscope, the attenuation value
cannot be changed from the sensed value. Attempting to set the oscilloscope to
an attenuation value other than the sensed value produces an error.
Query Syntax
:EXTernal:PROBe?
The :EXTernal:PROBe? query returns the current probe attenuation factor for the
external trigger.
Return Format
<attenuation><NL>
<attenuation> ::= probe attenuation ratio in NR3 format
See Also
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1428)
"Introduction to :EXTernal Trigger Commands"
":EXTernal:RANGe"
on page 386
"Introduction to :TRIGger Commands"
":CHANnel<n>:PROBe"
on page 1139
on page 311
:EXTernal Trigger Commands
on page 383
17
385

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the InfiniiVision MSO-X 4022A and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Infiniivision 4000 x seriesInfiniivision mso-x 4032aInfiniivision mso-x 4034aInfiniivision mso-x 4054aInfiniivision mso-x 4104aInfiniivision mso-x 4154a ... Show all

Table of Contents