Sbus<N>:Usb:trigger:data:length - Keysight Technologies InfiniiVision MSO-X 4022A Programmer's Manual

Hide thumbs Also See for InfiniiVision MSO-X 4022A:
Table of Contents

Advertisement

29
:SBUS<n> Commands
:SBUS<n>:USB:TRIGger:DATA:LENGth
Command Syntax
:SBUS<n>:USB:TRIGger:DATA:LENGth <length>
<length> ::= data length between 1-20
The :SBUS<n>:USB:TRIGger:DATA:LENGth command specifies the data length in
bytes.
Query Syntax
:SBUS<n>:USB:TRIGger:DATA:LENGth?
The :SBUS<n>:USB:TRIGger:DATA:LENGth? query returns the specified data
length.
Return Format
<length><NL>
<length> ::= data length between 1-20
See Also
980
(see
page
1428)
":SBUS<n>:USB:TRIGger:DATA"
on page 979
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide