Measure:bwidth - Keysight Technologies InfiniiVision MSO-X 4022A Programmer's Manual

Hide thumbs Also See for InfiniiVision MSO-X 4022A:
Table of Contents

Advertisement

:MEASure:BWIDth

Command Syntax
:MEASure:BWIDth [<source>]
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:BWIDth command installs a burst width measurement on screen. If
the optional source parameter is not specified, the current measurement source is
used.
This command is not available if the source is FFT (Fast Fourier Transform).
NOTE
Query Syntax
:MEASure:BWIDth? [<source>]
The :MEASure:BWIDth? query measures and returns the width of the burst on the
screen.
The burst width is calculated as follows:
Return Format
<value><NL>
<value> ::= burst width in seconds in NR3 format
See Also
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1428)
burst width = (last edge on screen - first edge on screen)
"Introduction to :MEASure Commands"
":MEASure:SOURce"
on page 529
:MEASure Commands
on page 488
22