Maintenance; Intervals - Weinmann MEDUMAT Easy CPR Instructions For Use Manual

Ventilators
Hide thumbs Also See for MEDUMAT Easy CPR:
Table of Contents

Advertisement

10 Maintenance

10.1 Intervals

Fault or treatment failure due to insufficient or incorrect
maintenance!
Incorrect maintenance can lead to dangerous situations and
failure or malfunctioning of the device. This can result in severe or
life-threatening injury to the patient and bystanders.
 Make sure that maintenance, safety checks and maintenance
measures are only performed by the manufacturer or
technicians who have been expressly authorized by the
manufacturer.
 Also observe and comply with the maintenance intervals for
devices in storage.
 Follow the maintenance intervals as stated on the device
labeling.
 Perform a full function check before every use.
Part concerned
Device
Reusable accessories
Testing bag
Fittings carrying oxygen
Mounting plate and the
fastening strap
Disposable hose system with
disposable patient valve
Interval
Maintenance by
Manufacturer or a technician
specifically authorized by the
Every 2 years
manufacturer
Customer Service", page 6)
Maintenance-free
MEDUMAT Easy
10 Maintenance
(see "1.2
CPR
EN
125

Advertisement

Table of Contents
loading

Table of Contents