5

Function check

5.1 Intervals for function check

5.2 Visually checking the device and
Requirement
If this function check reveals any faults or deviations from the
specified values, you must not use the MEDUMAT Easy
You should first try to rectify the fault with the aid of the
information provided in the section "Error messages"
Faults", page
123). If you are unable to rectify the faults using the
table, please contact WEINMANN Emergency or a technician who
has been expressly authorized by WEINMANN Emergency
promptly.
Devices and accessories which are defective or not ready for
use can disrupt the therapy or cause it to fail completely!
The use of defective devices or accessories can cause the device to
malfunction. This can result in severe or life-threatening injury to
the patient and user.
 Perform a full function check before every use.
 Only use devices and accessories which have successfully
passed the function check.
Perform the function check at the following intervals:
Part concerned
Device and accessories
accessories
The device is switched off
page
84).
1. Check the device and accessories for external damage.
2. Carefully bend the MEDUtrigger cable and check for:
Interval
Before each use
After each cleaning and disinfection
After each disassembly
At least every 6 months (if not used)
(see "6.6 Switching the device off",
MEDUMAT Easy
5 Function check
CPR
.
(see "9.2
CPR
EN