6.1.2 Connecting the breathing system filter
Fault or treatment failure due to incompatibility of the device
with consumables, accessories or other medical devices!
Defective and unauthorized accessories can result in malfunctions,
increased electromagnetic interference emissions and reduced
electromagnetic immunity of the device, incorrect output values
and reduced ventilation performance. This can result in serious or
life-threatening injury to the patient.
 Only connect approved accessories.
Hypoventilation due to the use of additional breathing system
filters!
The dead space of the overall system increases due to the use of
additional breathing system filters (breathing system filter,
bacteria filter or combined breathing system bacteria filter).
Increased dead space can result in hypoventilation. This can result
in serious or life-threatening injury to the patient.
 Only use approved accessories.
 Increase in the dead space with ventilation with small tidal
volumes.
Increased breathing effort due to the use of additional
accessories!
The spontaneous respiration resistance of the overall system
increases due to the use of additional accessories such as a filter
(breathing system filter, bacteria filter or combined breathing
system bacteria filter). This can injure the patient.
 Only use approved accessories.
 Monitor increase in spontaneous respiratory resistance for the
patient.
Connecting the breathing system filter for mask
ventilation
1. Observe the instructions for use from the breathing system
filter manufacturer.
CPR
MEDUMAT Easy
6 Operation
EN
67