6.5 Muting alarms
6.5.1 Activating alarm mute function
When an alarm is emitted, you can suppress the alarm tone for a
maximum of 120 seconds. The exception to this is the supply
pressure alarm < 2.7 bar O
be muted.
If the cause of the alarm continues to exist after 120 seconds, a
new alarm is emitted.
If the error is remedied, the visual and audible alarms are
automatically reset.
Alarm muting also applies to new alarms which are emitted within
these 120 seconds. The visual alarm is always active when in
muted state.
Requirement
•
•
1. Press the alarm mute button.
Result
•
•
6.5.2 Deactivating alarm mute function
Requirement
•
•
The device is switched on.
An alarm is triggered.
The audible alarm is deactivated for 120 seconds.
The orange LED above the alarm mute button lights up.
The audible alarm is activated
mute function", page
The orange LED above the alarm mute button lights up.
. The supply pressure alarm can not
2
(see "6.5.1 Activating alarm
83).
MEDUMAT Easy
6 Operation
CPR
EN
83